Fluke ScopeMeter Product Datasheet
Page 2
...Advanced trigger modes The ScopeMeter 190 Series simplifies triggering with the highest resolution and detail. This means that 's standard on an analog scope • F ast display update rate for troubleshooting time relationships between a high-resolution color display (190C Series) or blackand-white ...color labels, measurements and warnings are clearly linked to specific waveforms. See dynamic signal behavior instantaneously The Digital Persistence mode (Fluke 190C) helps to analyze complex dynamic signals by on Fast Fourier Transformation (FFT) analysis as 3000 samples per input,...
...Advanced trigger modes The ScopeMeter 190 Series simplifies triggering with the highest resolution and detail. This means that 's standard on an analog scope • F ast display update rate for troubleshooting time relationships between a high-resolution color display (190C Series) or blackand-white ...color labels, measurements and warnings are clearly linked to specific waveforms. See dynamic signal behavior instantaneously The Digital Persistence mode (Fluke 190C) helps to analyze complex dynamic signals by on Fast Fourier Transformation (FFT) analysis as 3000 samples per input,...
Fluke ScopeMeter Product Datasheet
Page 3
... may not be . Connect-and-View™ triggering for a stable display. Store complete memory content of the ScopeMeter on your Fluke ScopeMeter will automatically recognize a power failure and store the waveform data preceding it is measured on each of 100 captured screens with individual... ground, measurements on the part of the replay button. All inputs are instantly recognized and settings adjusted for an instant, stable display Scope users know how frustrating it . With 100 x zoom, you can look at the smallest details, like individual power cycles. providing...
... may not be . Connect-and-View™ triggering for a stable display. Store complete memory content of the ScopeMeter on your Fluke ScopeMeter will automatically recognize a power failure and store the waveform data preceding it is measured on each of 100 captured screens with individual... ground, measurements on the part of the replay button. All inputs are instantly recognized and settings adjusted for an instant, stable display Scope users know how frustrating it . With 100 x zoom, you can look at the smallest details, like individual power cycles. providing...
Fluke ScopeMeter Product Datasheet
Page 4
...or to the test point and letting the scope do the job right. scope, resistance and continuity measurements • 10:1 Voltage Probe included with Fluke 124 and 125 for more . With Connect-and-View, it comes with an oscilloscope. Fluke 125 Industrial ScopeMeter® Working under changing operational... you find errors like improper cable connections and terminators. The Fluke 125 validates the quality of 20 MHz (Fluke 123) or 40 MHz (Fluke 124, 125) the Fluke 120 Series will capture and display almost any signal. Scope mode With bandwidth of the electrical signals on each input....
...or to the test point and letting the scope do the job right. scope, resistance and continuity measurements • 10:1 Voltage Probe included with Fluke 124 and 125 for more . With Connect-and-View, it comes with an oscilloscope. Fluke 125 Industrial ScopeMeter® Working under changing operational... you find errors like improper cable connections and terminators. The Fluke 125 validates the quality of 20 MHz (Fluke 123) or 40 MHz (Fluke 124, 125) the Fluke 120 Series will capture and display almost any signal. Scope mode With bandwidth of the electrical signals on each input....
Fluke ScopeMeter Product Datasheet
Page 5
...Fluke 199B Fluke 196B Fluke 192B Fluke 125 Fluke 124 Fluke 123 200 MHz 100 MHz 200 MHz 100 MHz 60 MHz 40 MHz 40 MHz 20 MHz 2.5 GS/s 1.0 GS/s 2.5 GS/s 1.0 GS/s 500 MS/s 25 MS/s 25 MS/s 25 MS/s (covered by real time sample rate) 2.5 GS/s 2.5 GS/s 1.25 GS/s 3000 points 512 points (min/max pairs) 2 scope...- Ordering information Fluke 199C Fluke 199C/S Fluke 196C Fluke 196C/S Fluke 199B Fluke 199B/S Fluke 196B Fluke 196B/S Fluke 192B Fluke 192B/S Fluke 125 Fluke 125/S Fluke 124 Fluke 124/S Fluke 123 Fluke 123/S SCC190 SCC120...
...Fluke 199B Fluke 196B Fluke 192B Fluke 125 Fluke 124 Fluke 123 200 MHz 100 MHz 200 MHz 100 MHz 60 MHz 40 MHz 40 MHz 20 MHz 2.5 GS/s 1.0 GS/s 2.5 GS/s 1.0 GS/s 500 MS/s 25 MS/s 25 MS/s 25 MS/s (covered by real time sample rate) 2.5 GS/s 2.5 GS/s 1.25 GS/s 3000 points 512 points (min/max pairs) 2 scope...- Ordering information Fluke 199C Fluke 199C/S Fluke 196C Fluke 196C/S Fluke 199B Fluke 199B/S Fluke 196B Fluke 196B/S Fluke 192B Fluke 192B/S Fluke 125 Fluke 125/S Fluke 124 Fluke 124/S Fluke 123 Fluke 123/S SCC190 SCC120...
Fluke 123 and 124 Scopemeter Datasheet
Page 1
...input TrendPlot™ recorder all covered by bad connections, dust, dirt, corrosion or simply broken wiring or connectors. 40 MHz FLUKE-124/003S Industrial ScopeMeter 40 MHz with Fluke 124) • 600 V CAT III safety certified • Optically isolated RS-232 interface • Rugged, compact case ctrical Sa... Tools Three-in-one affordable, easy-to-use instrument. These are instantly recognized and settings adjusted for an instant, stable display Scope users know how difficult triggering can be . No time wasted finding or swapping leads. Use TrendPlot™ to help lead you...
...input TrendPlot™ recorder all covered by bad connections, dust, dirt, corrosion or simply broken wiring or connectors. 40 MHz FLUKE-124/003S Industrial ScopeMeter 40 MHz with Fluke 124) • 600 V CAT III safety certified • Optically isolated RS-232 interface • Rugged, compact case ctrical Sa... Tools Three-in-one affordable, easy-to-use instrument. These are instantly recognized and settings adjusted for an instant, stable display Scope users know how difficult triggering can be . No time wasted finding or swapping leads. Use TrendPlot™ to help lead you...
FE 123 & 124 Users Manual
Page 6
Fluke 123/124 Users Manual Freezing the Screen...16 Holding a Stable Reading 16 Making Relative Measurements 17 Selecting Auto/Manual Ranges 18 Changing the Graphic Representation on the ... Charging the Rechargeable Battery Pack 44 Keeping Batteries in Optimal Condition 45 Replacing and Disposing of the Rechargeable Battery Pack 46 Using and Adjusting 10:1 Scope Probes 47 Calibrating the Test Tool 49 Parts and Accessories 49 3 Tips and Troubleshooting 53 ii
Fluke 123/124 Users Manual Freezing the Screen...16 Holding a Stable Reading 16 Making Relative Measurements 17 Selecting Auto/Manual Ranges 18 Changing the Graphic Representation on the ... Charging the Rechargeable Battery Pack 44 Keeping Batteries in Optimal Condition 45 Replacing and Disposing of the Rechargeable Battery Pack 46 Using and Adjusting 10:1 Scope Probes 47 Calibrating the Test Tool 49 Parts and Accessories 49 3 Tips and Troubleshooting 53 ii
FE 123 & 124 Users Manual
Page 19
... the blue arrow keys to normal mode. Press the blue 'ENTER' function key to confirm selection Press 'ENTER' until you to open the Scope menu. 1 Using The Test Tool Making Selections in a Menu Subsequently follow steps to to open a menu and to check the menu without changing... the setup of the test tool. Press the SCOPE MENU key to step through a menu without destroying your settings. Note Pressing the SCOPE MENU key a second time closes this menu and resumes normal measurement. Figure 1-4.
... the blue arrow keys to normal mode. Press the blue 'ENTER' function key to confirm selection Press 'ENTER' until you to open the Scope menu. 1 Using The Test Tool Making Selections in a Menu Subsequently follow steps to to open a menu and to check the menu without changing... the setup of the test tool. Press the SCOPE MENU key to step through a menu without destroying your settings. Note Pressing the SCOPE MENU key a second time closes this menu and resumes normal measurement. Figure 1-4.
FE 123 & 124 Users Manual
Page 21
In the next example, the screen displays "1.411" in large numbers and "-0.103" in smaller numbers. Figure 1-6. The Screen After an Auto Set 13 A scope trace gives a graphical representation of the waveform. The zero icon ( ) identifies the ground level of the waveform. - 1 Using The Test Tool Displaying an Unknown Signal ...
In the next example, the screen displays "1.411" in large numbers and "-0.103" in smaller numbers. Figure 1-6. The Screen After an Auto Set 13 A scope trace gives a graphical representation of the waveform. The zero icon ( ) identifies the ground level of the waveform. - 1 Using The Test Tool Displaying an Unknown Signal ...
FE 123 & 124 Users Manual
Page 28
Highlight SMOOTH. Open the SCOPE OPTIONS submenu. Fluke 123/124 Users Manual Smoothing the Waveform To smooth the waveform, do the following: Open the SCOPE INPUTS menu. Smoothing the Waveform Jump to suppress noise without smoothing are shown in Figure 1-11. 20 Figure 1-11. Accept waveform smooth. You can use waveform smooth to WAVEFORM MODE. Waveform samples with and without loss of bandwidth.
Highlight SMOOTH. Open the SCOPE OPTIONS submenu. Fluke 123/124 Users Manual Smoothing the Waveform To smooth the waveform, do the following: Open the SCOPE INPUTS menu. Smoothing the Waveform Jump to suppress noise without smoothing are shown in Figure 1-11. 20 Figure 1-11. Accept waveform smooth. You can use waveform smooth to WAVEFORM MODE. Waveform samples with and without loss of bandwidth.
FE 123 & 124 Users Manual
Page 31
... To catch single events, you can perform a single shot. (One time screen update.) To set up the test tool for a Single Shot. 23 Open the SCOPE OPTIONS submenu. Highlight SINGLE SHOT. (2x) Accept the set up for a single shot on the input A waveform, do the following : • Connect the probe to...the TrendPlot Reading To toggle the TrendPlot reading between MIN (minimum), MAX (maximum), and AVERAGE, do the following : Change MIN into AVG reading. Open the SCOPE INPUTS menu. Note that the date and time stamp now updates continuously to be measured. Change MAX into MAX reading.
... To catch single events, you can perform a single shot. (One time screen update.) To set up the test tool for a Single Shot. 23 Open the SCOPE OPTIONS submenu. Highlight SINGLE SHOT. (2x) Accept the set up for a single shot on the input A waveform, do the following : • Connect the probe to...the TrendPlot Reading To toggle the TrendPlot reading between MIN (minimum), MAX (maximum), and AVERAGE, do the following : Change MIN into AVG reading. Open the SCOPE INPUTS menu. Note that the date and time stamp now updates continuously to be measured. Change MAX into MAX reading.
FE 123 & 124 Users Manual
Page 33
... MODE. (2x) Start Recording. Observe that during recording no measurements are only displayed after is especially useful when you measure lower frequency waveforms. Open the SCOPE INPUTS menu. Recording Slow Signals over Longer Period of waveform activity and is pressed. (See figure 1-15.) 25 The waveform moves accross the screen from...
... MODE. (2x) Start Recording. Observe that during recording no measurements are only displayed after is especially useful when you measure lower frequency waveforms. Open the SCOPE INPUTS menu. Recording Slow Signals over Longer Period of waveform activity and is pressed. (See figure 1-15.) 25 The waveform moves accross the screen from...
FE 123 & 124 Users Manual
Page 34
Fluke 123/124 Users Manual Selecting AC-Coupling Use AC-coupling when you wish to observe a small AC signal that rides on left of the waveform area. 26 ... a DC signal. Reversing the Polarity of INPUT A). Select NORMAL (of the Displayed Waveform To invert the input A waveform, do the following: Open the SCOPE INPUTS menu. Open the SCOPE INPUTS menu. Highlight INVERT. (3x) Accept inverted waveform display. Highlight AC. (4x) Accept AC-coupling for INPUT A. An inverted display is displayed as...
Fluke 123/124 Users Manual Selecting AC-Coupling Use AC-coupling when you wish to observe a small AC signal that rides on left of the waveform area. 26 ... a DC signal. Reversing the Polarity of INPUT A). Select NORMAL (of the Displayed Waveform To invert the input A waveform, do the following: Open the SCOPE INPUTS menu. Open the SCOPE INPUTS menu. Highlight INVERT. (3x) Accept inverted waveform display. Highlight AC. (4x) Accept AC-coupling for INPUT A. An inverted display is displayed as...
FE 123 & 124 Users Manual
Page 36
... text on the input A waveform, with automatic screen update, and to configure the auto range triggering for waveforms from 1 Hz, do the following: Open the SCOPE INPUTS menu. Highlight >1 Hz 28 Fluke 123/124 Users Manual Selecting the Trigger Parameters To trigger on bottom of the screen when no trigger is not valid.
... text on the input A waveform, with automatic screen update, and to configure the auto range triggering for waveforms from 1 Hz, do the following: Open the SCOPE INPUTS menu. Highlight >1 Hz 28 Fluke 123/124 Users Manual Selecting the Trigger Parameters To trigger on bottom of the screen when no trigger is not valid.
FE 123 & 124 Users Manual
Page 43
Figure 1-19. Voltage Measurement with Cursors 35 Making Cursor Measurements Fluke 124 has cursors. Press to the desired position on the screen. 1 Using The Test Tool Making Cursor Measurements Note Even when the key labels are displayed. ... on live waveforms and on saved waveforms. Using Horizontal Cursors on waveforms. This can still use the cursors for a voltage measurement, do the following: From Scope mode, display the Cursor Key functions. Highlight the lower cursor. } Move the lower cursor to make precise digital measurements on a Waveform To use the arrow...
Figure 1-19. Voltage Measurement with Cursors 35 Making Cursor Measurements Fluke 124 has cursors. Press to the desired position on the screen. 1 Using The Test Tool Making Cursor Measurements Note Even when the key labels are displayed. ... on live waveforms and on saved waveforms. Using Horizontal Cursors on waveforms. This can still use the cursors for a voltage measurement, do the following: From Scope mode, display the Cursor Key functions. Highlight the lower cursor. } Move the lower cursor to make precise digital measurements on a Waveform To use the arrow...
FE 123 & 124 Users Manual
Page 44
... desired position on the waveform. 36 } Highlight the right cursor. ~ Move the right cursor to highlight . If necessary, choose the trace A or B. Fluke 123/124 Users Manual The readout shows the voltage difference between the two cursors and the voltages at the cursors in relation to measure the amplitude, high... waveform. Time Measurement with Cursors Press to the desired position on a Waveform To use the cursors for a time measurement, do the following: From Scope mode, display the Cursor Key functions. Observe that two vertical cursors are displayed. See Figure 1-19).
... desired position on the waveform. 36 } Highlight the right cursor. ~ Move the right cursor to highlight . If necessary, choose the trace A or B. Fluke 123/124 Users Manual The readout shows the voltage difference between the two cursors and the voltages at the cursors in relation to measure the amplitude, high... waveform. Time Measurement with Cursors Press to the desired position on a Waveform To use the cursors for a time measurement, do the following: From Scope mode, display the Cursor Key functions. Observe that two vertical cursors are displayed. See Figure 1-19).
FE 123 & 124 Users Manual
Page 45
Making Rise Time Measurements To measure rise time, do the following: From Scope mode, display the Cursor Key functions. A marker is between the two markers (See Figure 1-20). Observe that two horizontal cursors are displayed. Move the upper ...
Making Rise Time Measurements To measure rise time, do the following: From Scope mode, display the Cursor Key functions. A marker is between the two markers (See Figure 1-20). Observe that two horizontal cursors are displayed. Move the upper ...
FE 123 & 124 Users Manual
Page 46
See Figure 1-21. ~ Turn off the cursors. 38 Using the 10:1 Probe for a Probe connected to the zero icon (-). Fluke 124 is much lower than that of this attenuation. The loading of the circuit by a 10:1 Probe is supplied with a high impedance. Open the Probes ...The Probe attenuates the signal 10 times. The example below is recommended when you measure high frequency signals in relation to input B: Open the Scope menu. Fluke 123/124 Users Manual The reading now shows the risetime from 10%-90% of the trace amplitude and the voltage at the cursors in circuits with...
See Figure 1-21. ~ Turn off the cursors. 38 Using the 10:1 Probe for a Probe connected to the zero icon (-). Fluke 124 is much lower than that of this attenuation. The loading of the circuit by a 10:1 Probe is supplied with a high impedance. Open the Probes ...The Probe attenuates the signal 10 times. The example below is recommended when you measure high frequency signals in relation to input B: Open the Scope menu. Fluke 123/124 Users Manual The reading now shows the risetime from 10%-90% of the trace amplitude and the voltage at the cursors in circuits with...
FE 123 & 124 Users Manual
Page 47
... these Probes is explained in the voltage readout. } Press ENTER to its inputs: high frequency adjustment is compensated in Chapter 2 under 'Using and Adjusting 10:1 Scope Probes'. 1 Using The Test Tool Using a Printer Using a Printer To print a (graphic) hard copy of the present screen, you need to use one of the...
... these Probes is explained in the voltage readout. } Press ENTER to its inputs: high frequency adjustment is compensated in Chapter 2 under 'Using and Adjusting 10:1 Scope Probes'. 1 Using The Test Tool Using a Printer Using a Printer To print a (graphic) hard copy of the present screen, you need to use one of the...
FE 123 & 124 Users Manual
Page 55
For Fluke 124 it is recommended to the input of the test tool. Warning To avoid electrical shock use the BB120 Banana-to-BNC adapter (delivered with the test tool) to connect a 10:1 scope probe to use the Fluke BP120 Ni-Cd (standard) or BP130 Ni-MH (extended operating time) ...2-3. For Fluke 123 use the Fluke BP130 Ni-MH battery pack . 8. Using and Adjusting 10:1 Scope Probes Note The 10:1 voltage probe that the battery pack is always adjusted correctly to the red input A jack. VPS100, optional) for optimal response. Note Ensure that is supplied with Fluke 124 is placed ...
For Fluke 124 it is recommended to the input of the test tool. Warning To avoid electrical shock use the BB120 Banana-to-BNC adapter (delivered with the test tool) to connect a 10:1 scope probe to use the Fluke BP120 Ni-Cd (standard) or BP130 Ni-MH (extended operating time) ...2-3. For Fluke 123 use the Fluke BP130 Ni-MH battery pack . 8. Using and Adjusting 10:1 Scope Probes Note The 10:1 voltage probe that the battery pack is always adjusted correctly to the red input A jack. VPS100, optional) for optimal response. Note Ensure that is supplied with Fluke 124 is placed ...
FE 123 & 124 Users Manual
Page 56
Open the PROBES submenu. Highlight PROBE AC ADJUST. A square wave appears on the screen. 48 Adjust the trimmer screw in the probe housing to normal mode. Open the PROBE AC ADJUST submenu. Return to give an optimum square wave. Fluke 123/124 Users Manual Open the SCOPE INPUTS menu. Highlight ADJUST 10:1 PROBE.
Open the PROBES submenu. Highlight PROBE AC ADJUST. A square wave appears on the screen. 48 Adjust the trimmer screw in the probe housing to normal mode. Open the PROBE AC ADJUST submenu. Return to give an optimum square wave. Fluke 123/124 Users Manual Open the SCOPE INPUTS menu. Highlight ADJUST 10:1 PROBE.