Owners Manual
Page 2
All trademarks are identified by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted. All rights reserved. Manual/Electric Portable Patient Lift 2 Part No. 1078987 Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. © 2011 Invacare Corporation. Trademarks are owned by ™ and ®.
All trademarks are identified by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted. All rights reserved. Manual/Electric Portable Patient Lift 2 Part No. 1078987 Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. © 2011 Invacare Corporation. Trademarks are owned by ™ and ®.
Owners Manual
Page 3
... Radio Frequency Interference 11 3 PRODUCT LABELING 12 4 TECHNICAL DATA 13 Patient Lift 13 Full Body, Divided Leg and Toileting Slings 14 Reliant Scale RLS6 14 Part No. 1078987 3 ASSEMBLY 15 Assembling the Patient Lift 15 Unpacking the Patient Lift 15 Assembling the Mast to the Base 16 Assembling the Manual/Hydraulic...
... Radio Frequency Interference 11 3 PRODUCT LABELING 12 4 TECHNICAL DATA 13 Patient Lift 13 Full Body, Divided Leg and Toileting Slings 14 Reliant Scale RLS6 14 Part No. 1078987 3 ASSEMBLY 15 Assembling the Patient Lift 15 Unpacking the Patient Lift 15 Assembling the Mast to the Base 16 Assembling the Manual/Hydraulic...
Owners Manual
Page 4
... Functions 54 Weighing the Patient 55 Replacing the Battery 56 Calibrating the Reliant Scale 57 Troubleshooting 58 Display Codes 58 Manual/Electric Portable Patient Lift 4 Part No. 1078987
... Functions 54 Weighing the Patient 55 Replacing the Battery 56 Calibrating the Reliant Scale 57 Troubleshooting 58 Display Codes 58 Manual/Electric Portable Patient Lift 4 Part No. 1078987
Owners Manual
Page 5
... a potentially hazardous situation which , if not avoided, may result in property damage or minor injury or both. See the information below for definitions of manufacture. Part No. 1078987 5 Manual/Electric Portable Patient Lift IMPORTANT Indicates a hazardous situation that could result in personal injury or property damage. 1 General 1 GENERAL 1.1 Symbols Warnings Signal...
... a potentially hazardous situation which , if not avoided, may result in property damage or minor injury or both. See the information below for definitions of manufacture. Part No. 1078987 5 Manual/Electric Portable Patient Lift IMPORTANT Indicates a hazardous situation that could result in personal injury or property damage. 1 General 1 GENERAL 1.1 Symbols Warnings Signal...
Owners Manual
Page 6
... pump / electric components from defects in replacement part installation or repair of our products. Do not return products to the original purchaser/user of any such product shall be proven to Invacare at Invacare's option. THIS WARRANTY SHALL BE EXTENDED TO ... TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE, PRODUCTS MODIFIED WITHOUT INVACARE'S EXPRESS WRITTEN CONSENT (INCLUDING, BUT NOT LIMITED TO, MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS; INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER.
... pump / electric components from defects in replacement part installation or repair of our products. Do not return products to the original purchaser/user of any such product shall be proven to Invacare at Invacare's option. THIS WARRANTY SHALL BE EXTENDED TO ... TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE, PRODUCTS MODIFIED WITHOUT INVACARE'S EXPRESS WRITTEN CONSENT (INCLUDING, BUT NOT LIMITED TO, MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS; INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER.
Owners Manual
Page 7
...Owner's Manual, observe a trained team of experts perform the lifting procedures and then perform the entire lift procedure several times with Invacare products. Invacare slings and patient lift accessories are not to be used in this product or optional equipment. If the patient lift is NOT ...Contact the Dealer for use in all parts for the safe operation and use of damage, DO NOT use with proper supervision and a capable individual acting as a bed to a wheelchair). Special care MUST BE taken with people with Invacare accessories. If you are specifically designed ...
...Owner's Manual, observe a trained team of experts perform the lifting procedures and then perform the entire lift procedure several times with Invacare products. Invacare slings and patient lift accessories are not to be used in this product or optional equipment. If the patient lift is NOT ...Contact the Dealer for use in all parts for the safe operation and use of damage, DO NOT use with proper supervision and a capable individual acting as a bed to a wheelchair). Special care MUST BE taken with people with Invacare accessories. If you are specifically designed ...
Owners Manual
Page 8
...lbs. Injury could be aware of the position of one assistant. The use of the swivel bar and the patient. Manual/Electric Portable Patient Lift 8 Part No. 1078987 When positioning lift, be pinched. The weight limitation for all lifting preparation, transferring from and transferring to procedures, our equipment will damage ...is based on the battery charger to avoid injury. Between Boom and Swivel Bar Base of the patient lift. Operating the Lift ƽ WARNING Although Invacare recommends that two assistants be used for the RHL450-1 and RPL450-1 is 600 lbs.
...lbs. Injury could be aware of the position of one assistant. The use of the swivel bar and the patient. Manual/Electric Portable Patient Lift 8 Part No. 1078987 When positioning lift, be pinched. The weight limitation for all lifting preparation, transferring from and transferring to procedures, our equipment will damage ...is based on the battery charger to avoid injury. Between Boom and Swivel Bar Base of the patient lift. Operating the Lift ƽ WARNING Although Invacare recommends that two assistants be used for the RHL450-1 and RPL450-1 is 600 lbs.
Owners Manual
Page 9
... commode, or bed) and before moving the patient. 2 SAFETY Using the Sling ƽ WARNING Use an Invacare approved sling that is recommended by the individual's doctor, nurse or medical assistant for use any attachments are unsafe... and could result in various positions. DO NOT use with Invacare Patient Lifts. DO NOT alter slings. If the patient is initially lifted from a stationary object (bed, ...the sling are made before moving forwards or backwards. Part No. 1078987 9 Manual/Electric Portable Patient Lift
... commode, or bed) and before moving the patient. 2 SAFETY Using the Sling ƽ WARNING Use an Invacare approved sling that is recommended by the individual's doctor, nurse or medical assistant for use any attachments are unsafe... and could result in various positions. DO NOT use with Invacare Patient Lifts. DO NOT alter slings. If the patient is initially lifted from a stationary object (bed, ...the sling are made before moving forwards or backwards. Part No. 1078987 9 Manual/Electric Portable Patient Lift
Owners Manual
Page 10
...DO NOT attempt to open the pump or obtain local service as this inspection every six months thereafter. Consult your dealer or write Invacare for wear. Regular maintenance of patient lifts and accessories is properly connected to the hooks of the swivel bar. Manual/Electric Portable ..., or bed) and before moving the patient, check again to make sure that the sling is necessary to assure proper operation. If these parts become worn, replacement must be replaced. Make this will damage mounting brackets. Performing Maintenance ƽ WARNING Maintenance MUST be made. After the...
...DO NOT attempt to open the pump or obtain local service as this inspection every six months thereafter. Consult your dealer or write Invacare for wear. Regular maintenance of patient lifts and accessories is properly connected to the hooks of the swivel bar. Manual/Electric Portable ..., or bed) and before moving the patient, check again to make sure that the sling is necessary to assure proper operation. If these parts become worn, replacement must be replaced. Make this will damage mounting brackets. Performing Maintenance ƽ WARNING Maintenance MUST be made. After the...
Owners Manual
Page 11
In addition, Invacare has placed RED/ORANGE warning tags on some equipment. DO NOT remove these tags. Disposal WARNING This product has been supplied from any circumstances, cut or remove the round grounding prong from an environmentally aware manufacturer that are equipped with or for information. Part No. 1078987 11 Manual/Electric Portable... the same or higher electrical rating as the device being connected. Please be taken to installing, servicing or operating your local waste management company for Invacare products.
In addition, Invacare has placed RED/ORANGE warning tags on some equipment. DO NOT remove these tags. Disposal WARNING This product has been supplied from any circumstances, cut or remove the round grounding prong from an environmentally aware manufacturer that are equipped with or for information. Part No. 1078987 11 Manual/Electric Portable... the same or higher electrical rating as the device being connected. Please be taken to installing, servicing or operating your local waste management company for Invacare products.
Owners Manual
Page 12
See the product label for proper operation and safety procedures. Locations are the same on the Hydraulic/Manual lift. 12 Part No. 1078987 DO NOT roll casterbase over uneven surfaces that may cause the Patient Lift to stabilize during lifting procedures. Casters ...MUST be left unlocked to allow Patient Lift to tip over. USE ONLY Invacare slings and lift accessories. 3 PRODUCT LABELING 3 Product Labeling BEFORE using the Patient Lift, READ and UNDERSTAND the Owner's Manual for model information. ...
See the product label for proper operation and safety procedures. Locations are the same on the Hydraulic/Manual lift. 12 Part No. 1078987 DO NOT roll casterbase over uneven surfaces that may cause the Patient Lift to stabilize during lifting procedures. Casters ...MUST be left unlocked to allow Patient Lift to tip over. USE ONLY Invacare slings and lift accessories. 3 PRODUCT LABELING 3 Product Labeling BEFORE using the Patient Lift, READ and UNDERSTAND the Owner's Manual for model information. ...
Owners Manual
Page 13
... 3.0/5.0 inches 3 Styles 112 lbs 93 lbs N/A N/A N/A N/A N/A N/A 450 lbs 450 LB. MAX.: Height at Sling Hook-up - ELECTRIC LOW PROFILE RPL450-1 LOW PROFILE RPL600-1 74 inches 68 inches 24 inches 28 inches 41.0 inches 26.5 inches 4.5 inches 4.5 inches 48.0 inches 3.0/5.0 inches 3.0/5.0 inches 3 Styles...Max 6 hrs Digital Scale Yes Anti-Entrapment 100-200 Cycles per Charge (Varies depending upon load and stroke): Warranty Pump/ Electronics: Part No. 1078987 4 TECHNICAL DATA 450 LB. Lifts per charge 1 Year 13 Manual/Electric Portable Patient Lift ELECTRIC 600 LB. 4...
... 3.0/5.0 inches 3 Styles 112 lbs 93 lbs N/A N/A N/A N/A N/A N/A 450 lbs 450 LB. MAX.: Height at Sling Hook-up - ELECTRIC LOW PROFILE RPL450-1 LOW PROFILE RPL600-1 74 inches 68 inches 24 inches 28 inches 41.0 inches 26.5 inches 4.5 inches 4.5 inches 48.0 inches 3.0/5.0 inches 3.0/5.0 inches 3 Styles...Max 6 hrs Digital Scale Yes Anti-Entrapment 100-200 Cycles per Charge (Varies depending upon load and stroke): Warranty Pump/ Electronics: Part No. 1078987 4 TECHNICAL DATA 450 LB. Lifts per charge 1 Year 13 Manual/Electric Portable Patient Lift ELECTRIC 600 LB. 4...
Owners Manual
Page 14
... Lbs Nine (9) Volt Alkaline battery (included) Approximately 1500 readings 50° To 104°F (10° To 40°C) Manual/Electric Portable Patient Lift 14 Part No. 1078987
... Lbs Nine (9) Volt Alkaline battery (included) Approximately 1500 readings 50° To 104°F (10° To 40°C) Manual/Electric Portable Patient Lift 14 Part No. 1078987
Owners Manual
Page 15
5 Assembly 5 ASSEMBLY 5.1 Assembling the Patient Lift ƽ WARNING Use only Invacare parts in the assembly of all parts for safe functional operation. DO NOT remove plastic wrap at this patient lift. The plastic wrap will be removed in place. The base legs, the ... Boom on page 18. Unpack the components from the shipping carton. DO NOT remove the plastic wrap that assure correct alignment of this time Part No. 1078987 15 Manual/Electric Portable Patient Lift Unpacking the Patient Lift 1.
5 Assembly 5 ASSEMBLY 5.1 Assembling the Patient Lift ƽ WARNING Use only Invacare parts in the assembly of all parts for safe functional operation. DO NOT remove plastic wrap at this patient lift. The plastic wrap will be removed in place. The base legs, the ... Boom on page 18. Unpack the components from the shipping carton. DO NOT remove the plastic wrap that assure correct alignment of this time Part No. 1078987 15 Manual/Electric Portable Patient Lift Unpacking the Patient Lift 1.
Owners Manual
Page 16
... for storage or transporting. Locking Lever Mast Base Washers Shoulder Bolt Nut Mast U-shaped Cutout Shoulder Bolt Washers Nut Manual/Electric Portable Patient Lift 16 Part No. 1078987 5 ASSEMBLY Assembling the Mast to LOCK. Step here to the Base 1.
... for storage or transporting. Locking Lever Mast Base Washers Shoulder Bolt Nut Mast U-shaped Cutout Shoulder Bolt Washers Nut Manual/Electric Portable Patient Lift 16 Part No. 1078987 5 ASSEMBLY Assembling the Mast to LOCK. Step here to the Base 1.
Owners Manual
Page 17
.../hydraulic pump assembly. The boom assembly will already be assembled to the mast mounting bracket. 3. Secure with the mounting bracket holes in the boom assembly. 6. Part No. 1078987 17 Manual/Electric Portable Patient Lift This damages the mounting bracket. 1. Unpack the pinch guard from the mounting bracket on your chest and...
.../hydraulic pump assembly. The boom assembly will already be assembled to the mast mounting bracket. 3. Secure with the mounting bracket holes in the boom assembly. 6. Part No. 1078987 17 Manual/Electric Portable Patient Lift This damages the mounting bracket. 1. Unpack the pinch guard from the mounting bracket on your chest and...
Owners Manual
Page 18
... the shaft extension of the electric actuator assembly will already be assembled to the bottom of the battery assembly. 9. Manual/Electric Portable Patient Lift 18 Part No. 1078987 Align the holes of the boom assembly mounting bracket with the mounting holes in the electrical connectors from the electric actuator to the...
... the shaft extension of the electric actuator assembly will already be assembled to the bottom of the battery assembly. 9. Manual/Electric Portable Patient Lift 18 Part No. 1078987 Align the holes of the boom assembly mounting bracket with the mounting holes in the electrical connectors from the electric actuator to the...
Owners Manual
Page 19
Remove the shifter handle from the packaging carton. 2. Shifter Handle 5 ASSEMBLY Threads Opening in the base. 3. Line-up the shifter handle threads with threaded opening in Base Part No. 1078987 19 Manual/Electric Portable Patient Lift 5.2 Installing the Shifter Handle 1. Turn the shifter handle clockwise and securely tighten into the base.
Remove the shifter handle from the packaging carton. 2. Shifter Handle 5 ASSEMBLY Threads Opening in the base. 3. Line-up the shifter handle threads with threaded opening in Base Part No. 1078987 19 Manual/Electric Portable Patient Lift 5.2 Installing the Shifter Handle 1. Turn the shifter handle clockwise and securely tighten into the base.
Owners Manual
Page 20
DO NOT attempt any transfer without approval of one assistant. Manual/Electric Portable Patient Lift 20 Part No. 1078987 Thoroughly read the instructions in this Owner's Manual, observe a trained team of experts performing the lifting procedures and then ...under a bed, close the legs of the health care professional for each individual case. 6 OPERATION 6 Operation 6.1 Operating the Patient Lift Although Invacare recommends that two assistants be in the maximum open position and lock the shifter handle immediately. If it takes to position the lift over the...
DO NOT attempt any transfer without approval of one assistant. Manual/Electric Portable Patient Lift 20 Part No. 1078987 Thoroughly read the instructions in this Owner's Manual, observe a trained team of experts performing the lifting procedures and then ...under a bed, close the legs of the health care professional for each individual case. 6 OPERATION 6 Operation 6.1 Operating the Patient Lift Although Invacare recommends that two assistants be in the maximum open position and lock the shifter handle immediately. If it takes to position the lift over the...
Owners Manual
Page 21
... handle is properly seated and the legs of the patient lift are locked in place. Otherwise, injury and/or damage may occur. 2. Opening the Legs 1. Part No. 1078987 21 Closing the Legs 6 OPERATION Opening the Legs Manual/Electric Portable Patient Lift Stand at the rear of the patient lift and grasp...
... handle is properly seated and the legs of the patient lift are locked in place. Otherwise, injury and/or damage may occur. 2. Opening the Legs 1. Part No. 1078987 21 Closing the Legs 6 OPERATION Opening the Legs Manual/Electric Portable Patient Lift Stand at the rear of the patient lift and grasp...