Fluke 375 FC CAL Support and Manuals
Get Help and Manuals for this Fluke item
This item is in your list!
View All Support Options Below
Free Fluke 375 FC CAL manuals!
Problems with Fluke 375 FC CAL?
Ask a Question
Free Fluke 375 FC CAL manuals!
Problems with Fluke 375 FC CAL?
Ask a Question
Popular Fluke 375 FC CAL Manual Pages
Product Manual - Page 1
MyFlukeStore www.
Shop for Fluke products online at:
.com 1.888.610.7664 All rights reserved.
All product names are trademarks of their respective companies. 374 FC/375 FC/376 FC/ 902 FC
Clamp Meter
Calibration Manual
February 2017 Rev. 1, 9/19
©2017-2019 Fluke Corporation.
Product Manual - Page 3
...902 FC Electrical Specifications ...6 Mechanical Specifications 7 Environmental Specifications 7 Performance Tests ...8 Calibration Adjustment ...9 Calibration Setup ...11 VAC Adjustment Procedure 12 VDC/mVDC Adjustment Procedure 13 Ohm/Cap Adjustment Procedure 13 AAC Adjustment Procedure (374 FC, 375 FC, 376 FC 14 AAC Adjustment Procedure (902 FC 14 ADC Adjustment Procedure (374 FC, 375 FC...
Product Manual - Page 4
... provides all the information necessary to the 375 FC/375 FC/376 FC Users Manual and the 902 FC Users Manual
1
MyFlukeStore Shop for the 374 FC/375 FC/376 FC and the 902 FC Clamp Meters (the Product).
For complete operating instructions, refer to perform basic maintenance and make calibration adjustments for Fluke products online at: www.
.com 1.888.610.7664 Introduction
XW Warning...
Product Manual - Page 5
... injury:
• Carefully read all instructions.
• Read all probes, test...door is clean, dry, and in good repair. Examine the test leads for damaged insulation and...codes. Also look for Fluke products online at: www.
.com 1.888.610.7664 374 FC/375 FC/376 FC and 902 FC Calibration Manual...Product behind the tactile barrier.
• Replace the batteries when the low battery indicator ...
Product Manual - Page 6
... damaged, the Product may not be safe to use the HOLD function to Fluke for replacement of test probe, The protective cap decreases the expose probe metal Return the Product... input signals before you clean the Product. • Use only specified replacement parts. • When batteries are changed, ensure that the calibration seal in CAT III or CAT IV environments without the protective cap of...
Product Manual - Page 7
... MAINS installation. The Product
374 FC/375 FC/376 FC Electrical Specifications
AC Current via Jaw
Range 374 FC and 375 FC 600.0 A 376 FC 999.9 A
Resolution 0.1 A Accuracy 2 % ±5 digits (10 Hz to100 Hz)
2.5 % ±5 digits (100 to 500 Hz) Crest Factor (50 Hz/60 Hz)
376 FC 3 @ 500 A 2.5 @ 600 A 1.42 @ 1000 A
374 FC and 375 FC 2.5 @ 350 A 1.42 @ 600 A
Note: Add 2 % for Fluke products...
Product Manual - Page 9
374 FC/375 FC/376 FC and 902 FC Calibration Manual
Frequency via Jaw
Range 375 FC and 376 FC 5.0 Hz to 500.0 Hz
Resolution 0.1 Hz Accuracy 0.5 % ±5 digits
Trigger Level 5 Hz to 10 Hz, ≥10 A 10 Hz to 100 Hz, ≥5 A 100 Hz to 500 Hz, ≥10 A
Frequency via Flexible Current Probe
Range 375 FC and 376 FC 5.0 Hz to 500...
Product Manual - Page 11
...Calibrator. 3. Set the UUT to Table 2. (Where X = not applicable.)
The test passes if the UUT reads between the upper and lower limit.
Do the performance tests periodically and after service or repair. See Calibration Adjustment. 374 FC/375 FC/376 FC and 902 FC Calibration Manual...102.5 612.5 995
MyFlukeStore 8 Shop for Fluke products online at: www.
.com 1.888.610.7664 Table ...
Product Manual - Page 12
...Model
Fluke 55xxA Calibrator 5500A/COIL PN 666602 PN 2070140 Common power supply or a 2 x AA or AAA battery container
MyFlukeStore Shop for the calibration adjustment. Performance Tests (cont.)
Function
Calibrator...;A 200 μA -200 μA
-40 °C 0 °C 50 °C 400 °C
374 FC 375 FC
X
X
√
√
√
√
√
√
√
√
√
√...
Product Manual - Page 13
...; Never touch open-edge connectors except at : www.
.com 1.888.610.7664 Learn the guidelines for Fluke products online at a
• Use grounded-tip soldering irons only.
parts only at a staticfree work station. 374 FC/375 FC/376 FC and 902 FC Calibration Manual
h Static Awareness h
X
Semiconductors and integrated circuits can be damaged by electrostatic discharge during handling.
Product Manual - Page 15
...all the steps, push to save the data to NVRAM storage. Replace the battery compartment door and tighten the battery compartment screw.
c. c. Current Calibration Setup
ghn50.eps
9. b. Wait until each calibration adjustment: a. 374 FC/375 FC/376 FC and 902 FC Calibration Manual
5522A CALIBRATOR
NORMAL V, , ,RTD
A,
AUX -SENSE, AUX V
HI
SCOPE OUT
LO TRIG
GUARD
20A...
Product Manual - Page 16
...C-16 C-17 C-18 C-19 C-20 C-21 C-22 C-23 Save
374 FC 375 FC
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
√
376 FC
902 FC
Calibrator Output
√
0 Ω
√
600 Ω
√
660 Ω
√...
Product Manual - Page 17
374 FC/375 FC/376 FC and 902 FC Calibration Manual
AAC Adjustment Procedure (374 FC, 375 FC, 376 FC)
Step
LCD Display
374 FC 375 FC
1
C-24
√
√
2
C-25
√
√
3
C-26
√
√
4
C-27
√
√
5
C-28
√
√
6
C-29
√
√
7
C-30
√
√
8
C-31
√
√
9
Save
...
Product Manual - Page 18
...LCD Display C-37 Save
902 FC
√
0 °C
√
STBY
Calibrator Output
Maintenance
This section contains basic maintenance procedures. Battery Replacement
XW Warning To prevent possible electrical shock, fire, or personal injury, and to
6. Clamp Meter Maintenance
ROGO Adjustment Procedure (374 FC, 375 FC, 376 FC)
Step
LCD Display
374 FC 375 FC
1
C-34
√
√...
Product Manual - Page 19
...374 FC/375 FC/376 FC and 902 FC Calibration Manual
Replacement Parts Replacement parts and accessories are listed in Tables 4. Table 4. Replacement Parts and Accessories
Fluke Part Number 4696918 3752958 1997276 4705494 4748982 376756 855742 3798105 1997234 4744076 4744083 4698317
Description Battery Door Assembly Soft Case (374 FC/375 FC/376 FC) Soft Case(902 FC) User Manual (374 FC/375 FC...
Fluke 375 FC CAL Reviews
Do you have an experience with the Fluke 375 FC CAL that you would like to share?
Earn 750 points for your review!
We have not received any reviews for Fluke yet.
Earn 750 points for your review!