Calibration Guide - Page 1

1551A Ex/1552A Ex Stik Thermometer Calibration Manual November 2010 © 2010 Fluke Corporation. All product names are subject to change without notice. All rights reserved. Printed in USA. Specifications are trademarks of their respective companies.
Calibration Guide - Page 2

... is one country is submitted for 90 days and that service center, with its functional specifications for repair in Fluke's opinion, has been misused, altered, neglected, contaminated, or damaged by a court or other provision. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one year and begins on new...
Calibration Guide - Page 3

... Fluke 1 Safety Information 2 ATEX Safety Information 3 Symbols ...3 Approvals ...4 Specifications...4 Calibration ...5 Overview ...5 Recommended Equipment 6 Initial Setup 6 Record as-found Calibration Parameters 7 Evaluate as-found Measurement Errors 7 Adjustment at the Low Calibration Point 8 Adjustment at the Zero Calibration Point 9 Adjustment at the High Calibration Point 9 Set...
Calibration Guide - Page 9

... at www.fluke.com. For complete usage instructions, see , print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals. 1 Introduction The Stik Thermometer (the Product) is a precision digital thermometer that can be used in many of the same hazardous locations as analog thermometers. To see the 1551A Ex/1552A Ex Users Manual. It...
Calibration Guide - Page 10

.... If you open the Product case. 1551A Ex/1552A Ex Calibration Manual Safety Information A Warning identifies conditions and actions...Replace the batteries when the low battery indicator shows to the manufacturer for inspection. • The Product is intended for installation...Product to the user. Refer to the battery replacement instructions for proper operation of data. Should the Product...
Calibration Guide - Page 11

... cause a safety risk. To make sure of safe Product operation, fully follow these instructions, personal injury or Product damage can occur. These areas are not sure (due to translation and/or printing errors), refer to the English manual. Table 1. See Manual. Substitution of Danger. Important information. X Hazardous Voltage Do not dispose of this product...
Calibration Guide - Page 12

...Manufactured by Fluke Corporation: Fluke Corporation P.O. 1551A Ex/1552A Ex Calibration Manual Approvals The Product meets the above requirements using: EN 60079-0:2006 ELECTRICAL APPARATUS FOR EXPLOSIVE GAS ATMOSPHERES -- P.O. PART 0: GENERAL REQUIREMENTS (IEC 60079-0:2004 (MOD)) EN 60079-11:2007 EXPLOSIVE ATMOSPHERES -- Box 1186 5602 BD Eindhoven The Netherlands Specifications (Ambient: 23...
Calibration Guide - Page 13

... as-left measurement errors • Record as-left calibration parameters values 5 Probe Size 1551A-9: 4.8 mm x 229 mm (3/16 in . x 20 in.) 1552A: 6.35 mm x 305 mm (1/4 in . Each laboratory should only be adjusted to 30 minutes or can be performed by an uncertainty analysis also based on their own specific procedure based on the...
Calibration Guide - Page 14

... in the Setup menu. 1551A Ex/1552A Ex Calibration Manual Recommended Equipment The equipment listed in the Product and replace them if necessary. The battery level indication is usually 9600 but may be set to communicate with the 1551A Ex/1552A Ex Thermometer Initial Setup Check the batteries in Table 2 will satisfy the calibration requirements: Table 2. Several settings in the...
Calibration Guide - Page 15

Stik Thermometer Calibration Record as -found Measurement Errors The Product should be recorded. As-found Calibration Parameters Before any adjustments are made , the present values of the bath, reference thermometer, and 1551A Ex/1552A Ex Thermometer are given in Table 4, depending on the model. Evaluate as -found Calibration Parameters Parameter Command Value (fill in) TEMP1 ADJ1...
Calibration Guide - Page 16

... Ex Calibration Manual At each calibration point the measurement error may be necessary to measure the given temperature accurately. Low-Adjustment Temperature Point Low −50 °C 1551A Ex −80 °C 1552A Ex Before adjustments to any calibration parameters can be enabled by sending the following command: CAL:USER:LOW The parameter is the current password...
Calibration Guide - Page 17

...1551A Ex 0 °C 0 °C 1552A Ex To make an automatic adjustment at the High calibration...Calibration Point The final adjustment is made for the high end of the temperature range of the Product. Set...This adjustment will set the value ... Calibration Adjustment at the Zero Calibration Point...1551A Ex 157 °C 300 °C 1552A Ex To make an automatic adjustment at the Zero calibration...
Calibration Guide - Page 18

... to a final test to measure the given temperature accurately. The measurements are collected and errors are evaluated in the same way that was explained in ) 10 Point 1 2 3 4 5 Table 9. 1551A Ex/1552A Ex Calibration Manual The command will set , protection of the calibration parameters should be restored using the following command: SYST:PASS:CDIS (Protection will always...
Calibration Guide - Page 19

...3. HANDLE S.S. DEVICES BY THE BODY. static awareness A Message From Fluke Corporation Some semiconductors and custom IC's can minimize the chances of destroying...following practices should be damaged by : 1. KEEP PARTS IN ORIGINAL CONTAINERS UNTIL READY FOR USE. 4. Using the procedures, packaging, and bench techniques that there is a problem. 2. This notice explains how you can be...
Calibration Guide - Page 22

... rating. 4. Battery Alkaline AAA Figure 1. Replace the battery door and install the screw. Remove the battery door to remove the battery door screw. 2. Battery Replacement Table 11. Use a flat-blade screwdriver... gjo003.eps 12 Replace only with approved AAA batteries listed in this document. 1551A Ex/1552A Ex Calibration Manual To change the batteries, see Figure 1: 1.

Fluke Calibration 1551A-12-DL Reviews

Do you have an experience with the Fluke Calibration 1551A-12-DL that you would like to share?
Earn 750 points for your review!
We have not received any reviews for Fluke yet.

Ask a New Question
Use the box below to post a new question about Fluke Calibration 1551A-12-DL.
Manuals / Documents
Download any of our Fluke Calibration 1551A-12-DL manuals for free!

We have the following 1 documents available for the Fluke Calibration 1551A-12-DL:
  • Calibration Guide

Points & Prizes
  • You can earn points for nearly everything you do on HelpOwl.com
  • You can trade in those points for gift cards at leading retailers such as Amazon.com and Walmart
  • It's that simple!
See How it Works
Create a Free Account

Fluke Manuals

Find free Fluke Calibration 1551A-12-DL manuals and user guides available at ManualOwl.com. Try out our unique manual viewer allowing you to interact with manuals from directly within your browser!

Fluke Reviews

View thousands of Fluke Calibration 1551A-12-DL user reviews and customer ratings available at ReviewOwl.com.

Contact Information

Complete Fluke customer service contact information including steps to reach representatives, hours of operation, customer support links and more from ContactHelp.com.

Scoreboard Ratings

See detailed Fluke customer service rankings, employee comments and much more from our sister site.

Fluke Recalls

Find comprehensive Fluke recall information updated hourly on RecallOwl.com.