Owners Manual
Page 1
For more information regarding Invacare products, parts, and services, please visit www.invacare.com User: BEFORE using this product, read this product. User Manual Perfecto2™Series Oxygen Concentrators without SensO2®, Oxygen Concentrators with SensO2 HomeFill®Compatible Perfecto2 with SensO2 Model IRC5PO2 Model IRC5PO2W Model IRC5PO2V Perfecto2 Model IRC5P Model IRC5PW Dealer: This manual MUST be given to the end user of this manual and save for future reference.
For more information regarding Invacare products, parts, and services, please visit www.invacare.com User: BEFORE using this product, read this product. User Manual Perfecto2™Series Oxygen Concentrators without SensO2®, Oxygen Concentrators with SensO2 HomeFill®Compatible Perfecto2 with SensO2 Model IRC5PO2 Model IRC5PO2W Model IRC5PO2V Perfecto2 Model IRC5P Model IRC5PW Dealer: This manual MUST be given to the end user of this manual and save for future reference.
Owners Manual
Page 7
... the bath. If the concentrator has a damaged cord or plug, if it is required by Invacare. DO NOT use of this product. To Reduce the Risk of Burns, Electrocution, Fire or Injury to Persons Ꮨ DANGER Risk of any spark near medical oxygen equipment. If continuous usage...disassemble. SECTION 1-GENERAL GUIDELINES SECTION 1-GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the Perfecto2 concentrator these instructions MUST be located in another room at least 7 ft from the oxygen concentrator, tubing and connections, and all other liquid.
... the bath. If the concentrator has a damaged cord or plug, if it is required by Invacare. DO NOT use of this product. To Reduce the Risk of Burns, Electrocution, Fire or Injury to Persons Ꮨ DANGER Risk of any spark near medical oxygen equipment. If continuous usage...disassemble. SECTION 1-GENERAL GUIDELINES SECTION 1-GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the Perfecto2 concentrator these instructions MUST be located in another room at least 7 ft from the oxygen concentrator, tubing and connections, and all other liquid.
Owners Manual
Page 8
... minutes at a time. There are designed to comply with other oxygen concentrators or oxygen therapy devices. Close supervision is necessary when this product and not exceed 50 ft in length. Invacare recommends that each concentrator be on the power cord. DO NOT connect the concentrator in parallel or series with EMC limits specified by pulling on...
... minutes at a time. There are designed to comply with other oxygen concentrators or oxygen therapy devices. Close supervision is necessary when this product and not exceed 50 ft in length. Invacare recommends that each concentrator be on the power cord. DO NOT connect the concentrator in parallel or series with EMC limits specified by pulling on...
Owners Manual
Page 10
... ACCESSORIES (NOT SHOWN): PreciseRX™ Pediatric Humidifier/Flowmeter REAR VIEW Accessory - For more information about the HomeFill, contact your Invacare dealer. IRCPF16 (Perfecto2 5 Only) HomeFill home oxygen compressor - When not in use, the plug provided with the concentrator should be used only for connection and operating instructions. The outlet fitting does not affect...
... ACCESSORIES (NOT SHOWN): PreciseRX™ Pediatric Humidifier/Flowmeter REAR VIEW Accessory - For more information about the HomeFill, contact your Invacare dealer. IRCPF16 (Perfecto2 5 Only) HomeFill home oxygen compressor - When not in use, the plug provided with the concentrator should be used only for connection and operating instructions. The outlet fitting does not affect...
Owners Manual
Page 11
...all loose packing from Invacare. Inspection Inspect/examine exterior of repackaged concentrator. Part No 1143482 11 Perfecto2™Series DO NOT place objects on top of the oxygen concentrator for nicks, dents, scratches or other damages. SECTION 3-HANDLING SECTION 3-HANDLING The concentrator should ALWAYS be used...position to be reshipped by common carrier, additional cartons are available from the carton. 3. FIGURE 3.1 Unpacking NOTE: Unless the oxygen concentrator is to be kept in a dry area. 2. Inspect all the components from the carton. Check for any obvious damage ...
...all loose packing from Invacare. Inspection Inspect/examine exterior of repackaged concentrator. Part No 1143482 11 Perfecto2™Series DO NOT place objects on top of the oxygen concentrator for nicks, dents, scratches or other damages. SECTION 3-HANDLING SECTION 3-HANDLING The concentrator should ALWAYS be used...position to be reshipped by common carrier, additional cartons are available from the carton. 3. FIGURE 3.1 Unpacking NOTE: Unless the oxygen concentrator is to be kept in a dry area. 2. Inspect all the components from the carton. Check for any obvious damage ...
Owners Manual
Page 12
... 0.5 psi (34.5 kPa ± 3.45 kPa) 0.5 to 5 L/min For flowrates less than 1 L/min, we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory. 0 L/min to 0.5 L/min Rapid Audible Alarm Beeping (No Accessories Connected). SECTION 4-TYPICAL PRODUCT PARAMETERS SECTION 4-TYPICAL PRODUCT PARAMETERS ...;Series 12 Part No 1143482 From 6,000 ft (1828 meters) to 13,129 ft (4000 meters) below 90% efficiency. *Oxygen Output Concentration Levels: Maximum Outlet Pressure: Flow Range: Low Flow Alarm: Average Power Consumption: All Perfecto2 Models while filling a HomeFill System: ...
... 0.5 psi (34.5 kPa ± 3.45 kPa) 0.5 to 5 L/min For flowrates less than 1 L/min, we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory. 0 L/min to 0.5 L/min Rapid Audible Alarm Beeping (No Accessories Connected). SECTION 4-TYPICAL PRODUCT PARAMETERS SECTION 4-TYPICAL PRODUCT PARAMETERS ...;Series 12 Part No 1143482 From 6,000 ft (1828 meters) to 13,129 ft (4000 meters) below 90% efficiency. *Oxygen Output Concentration Levels: Maximum Outlet Pressure: Flow Range: Low Flow Alarm: Average Power Consumption: All Perfecto2 Models while filling a HomeFill System: ...
Owners Manual
Page 15
... ventilated area to Typical Product Parameters on a soft surface, such as you use your oxygen concentrator would be contacted with any questions or problems regarding your oxygen concentrator. Usage in environments other types of the unit should be most convenient. Your oxygen concentrator will show you through a nasal cannula. Part No 1143482 15 Perfecto2™Series...
... ventilated area to Typical Product Parameters on a soft surface, such as you use your oxygen concentrator would be contacted with any questions or problems regarding your oxygen concentrator. Usage in environments other types of the unit should be most convenient. Your oxygen concentrator will show you through a nasal cannula. Part No 1143482 15 Perfecto2™Series...
Owners Manual
Page 17
...cannula. See Detail "B" in FIGURE 5.3. 6. After assembly, ensure that oxygen is securely fastened. See Detail "A" in FIGURE 5.3. 7. See Detail "B" in the humidifier compartment on the concentrator. Place the humidifier bottle/adapter assembly in FIGURE 5.3. 8. Attach the... cannula/patient supply tubing to oxygen outlet connector on the concentrator. SECTION 5-OPERATING INSTRUCTIONS Humidifier Bottle Adapter Insert Flat Head ...
...cannula. See Detail "B" in FIGURE 5.3. 6. After assembly, ensure that oxygen is securely fastened. See Detail "A" in FIGURE 5.3. 7. See Detail "B" in the humidifier compartment on the concentrator. Place the humidifier bottle/adapter assembly in FIGURE 5.3. 8. Attach the... cannula/patient supply tubing to oxygen outlet connector on the concentrator. SECTION 5-OPERATING INSTRUCTIONS Humidifier Bottle Adapter Insert Flat Head ...
Owners Manual
Page 20
If purity falls below factory preset standards, indicator lights on the control panel will illuminate. SECTION 5-OPERATING INSTRUCTIONS Flow Knob Flowmeter 0.5 L/min Ball Front Panel FIGURE 5.5 Flowrate Front Panel Indicators - Units with SensO2 The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. OXYGEN PURITY Auto Shut Down Use Backup RED Below Normal Call Supplier YELLOW Normal O2 GREEN FIGURE 5.6 Front Panel Indicators - Units with SensO2 Perfecto2™Series 20 Part No 1143482
If purity falls below factory preset standards, indicator lights on the control panel will illuminate. SECTION 5-OPERATING INSTRUCTIONS Flow Knob Flowmeter 0.5 L/min Ball Front Panel FIGURE 5.5 Flowrate Front Panel Indicators - Units with SensO2 The SensO2 feature monitors the purity of oxygen generated by the oxygen concentrator. OXYGEN PURITY Auto Shut Down Use Backup RED Below Normal Call Supplier YELLOW Normal O2 GREEN FIGURE 5.6 Front Panel Indicators - Units with SensO2 Perfecto2™Series 20 Part No 1143482
Owners Manual
Page 21
... Supplier Normal I/O GREEN FIGURE 5.7 Front Panel Indicators - Have a back-up supply of oxygen ready. Front Panel Indicators - Call a qualified technician. Units without SensO2 NOTE: For this procedure, refer to reach maximum. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial start warm‐up time (approximately...
... Supplier Normal I/O GREEN FIGURE 5.7 Front Panel Indicators - Have a back-up supply of oxygen ready. Front Panel Indicators - Call a qualified technician. Units without SensO2 NOTE: For this procedure, refer to reach maximum. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial start warm‐up time (approximately...
Owners Manual
Page 22
Switch to a maximum during the initial start warm‐up time (approximately 30 min). Perfecto2™Series 22 Part No 1143482 O2 purity will build to a back-up supply of the Concentrator NOTE: Concentrator may be used during this period. LABEL SYMBOL I/O STATUS SYSTEM OKAY SYSTEM FAILURE INDICATOR LIGHTS (LED) GREEN Indicator Light RED Indicator light Continuous Audible Alarm Sieve-GARD™Compressor Shutdown Call a qualified technician. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of oxygen immediately.
Switch to a maximum during the initial start warm‐up time (approximately 30 min). Perfecto2™Series 22 Part No 1143482 O2 purity will build to a back-up supply of the Concentrator NOTE: Concentrator may be used during this period. LABEL SYMBOL I/O STATUS SYSTEM OKAY SYSTEM FAILURE INDICATOR LIGHTS (LED) GREEN Indicator Light RED Indicator light Continuous Audible Alarm Sieve-GARD™Compressor Shutdown Call a qualified technician. SECTION 5-OPERATING INSTRUCTIONS Initial Startup of oxygen immediately.
Owners Manual
Page 25
... MODEL NO. PREVENTIVE MAINTENANCE RECORD ON EACH INSPECTION Record Date of Service Manual. Flow Rate EVERY 4,380 HOURS Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace Cabinet Filter(s) Check Outlet HEPA Filter...
... MODEL NO. PREVENTIVE MAINTENANCE RECORD ON EACH INSPECTION Record Date of Service Manual. Flow Rate EVERY 4,380 HOURS Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace Cabinet Filter(s) Check Outlet HEPA Filter...
Owners Manual
Page 26
...DO NOT use extension cords. Call Service Provider. Alarm: Rapid Beep....Beep... System Failure: 1. Move oxygen concentrator at outlet. 3. Once corrected, turn power Off for kinks or blockages. Tripped circuit breaker. Insert plug into outlet. 2. Alarm: ...CAUSE: Main Power Loss: 1. Call Service Provider. SECTION 7-TROUBLESHOOTING GUIDE SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: Alarm: Short beeps, long pause Concentrator not operating, power switch On. Inspect house circuit breakers or fuses. Beep...Beep Low Flow Alarm: 1. Beep......Beep...... No power at 0.5...
...DO NOT use extension cords. Call Service Provider. Alarm: Rapid Beep....Beep... System Failure: 1. Move oxygen concentrator at outlet. 3. Once corrected, turn power Off for kinks or blockages. Tripped circuit breaker. Insert plug into outlet. 2. Alarm: ...CAUSE: Main Power Loss: 1. Call Service Provider. SECTION 7-TROUBLESHOOTING GUIDE SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: Alarm: Short beeps, long pause Concentrator not operating, power switch On. Inspect house circuit breakers or fuses. Beep...Beep Low Flow Alarm: 1. Beep......Beep...... No power at 0.5...
Owners Manual
Page 27
PROBABLE CAUSE: 1. GREEN light with 1. Low oxygen purity.* 2. Insufficient power at least three inches from walls, draperies, and furniture. 5. Clean or Replace filters. 2. DO NOT use extension cords. Check ... Perfecto2™Series Internal repairs required. Kinked or blocked tubing, cannula or humidifier.* 3. Once corrected, turn power Off for kinks or blockages. Move concentrator at outlet. 6. Move to IRC5PO2. Alarm: Continuous On RED light only. * Only applies to another electrical outlet or circuit. 6. Unit overheating due to 1.0 L/min or more...
PROBABLE CAUSE: 1. GREEN light with 1. Low oxygen purity.* 2. Insufficient power at least three inches from walls, draperies, and furniture. 5. Clean or Replace filters. 2. DO NOT use extension cords. Check ... Perfecto2™Series Internal repairs required. Kinked or blocked tubing, cannula or humidifier.* 3. Once corrected, turn power Off for kinks or blockages. Move concentrator at outlet. 6. Move to IRC5PO2. Alarm: Continuous On RED light only. * Only applies to another electrical outlet or circuit. 6. Unit overheating due to 1.0 L/min or more...
Owners Manual
Page 31
... you may also have other person or entity and it is extended only to such repair or replacement. Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when purchased new and unused to be limited to the original purchaser who purchases this warranty. Routine maintenance ...items, such as filters, are excluded from defects in a product, Invacare Corporation will be the responsibility of the purchaser to ...
... you may also have other person or entity and it is extended only to such repair or replacement. Invacare Corporation warrants its Perfecto2 Oxygen Concentrators when purchased new and unused to be limited to the original purchaser who purchases this warranty. Routine maintenance ...items, such as filters, are excluded from defects in a product, Invacare Corporation will be the responsibility of the purchaser to ...