Owners Manual
Page 1
microAIR® MA65 Series MA65, MA65B42, MA65B48, MA65RSR en Alternating Pressure Low Air Loss Mattress User Manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
microAIR® MA65 Series MA65, MA65B42, MA65B48, MA65RSR en Alternating Pressure Low Air Loss Mattress User Manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
Owners Manual
Page 3
...5.1.3 Adjusting the Minimum Pressure Setting (A/P Low 20 5.1.4 Static Button 21 5.1.5 Firm/Soft Buttons 21 5.1.6 Fowler Button 21 5.1.7 Max Inflate/Low Air Loss Button 22 5.1.8 Lock/Alarm Silence Button 23 5.1.9 Power Fail LED 23 5.1.10 Low Pressure LED 23 5.2 Powering Up the System 23 5.3... Placing the Patient on the Mattress 24 5.4 Transferring Patient From/To a Gurney 25 5.5 Transferring Patient From/To a Wheelchair 25 5.6 Preparing for CPR Procedure 26 5.7 About Power...
...5.1.3 Adjusting the Minimum Pressure Setting (A/P Low 20 5.1.4 Static Button 21 5.1.5 Firm/Soft Buttons 21 5.1.6 Fowler Button 21 5.1.7 Max Inflate/Low Air Loss Button 22 5.1.8 Lock/Alarm Silence Button 23 5.1.9 Power Fail LED 23 5.1.10 Low Pressure LED 23 5.2 Powering Up the System 23 5.3... Placing the Patient on the Mattress 24 5.4 Transferring Patient From/To a Gurney 25 5.5 Transferring Patient From/To a Wheelchair 25 5.6 Preparing for CPR Procedure 26 5.7 About Power...
Owners Manual
Page 10
...(30,5 cm) away from lint, hair or similar foreign items. - Contact a qualified technician or Invacare if any adjustments, repair or service and before use in a well ventilated space so that the air intake and exhaust are present: - Parts showing excessive wear should be replaced. - After any of ...the bed frame at the center of Death, Injury, or Damage To avoid death, injury or damage due to the mattress: - Risk of the bed. microAIR® MA65 Series WARNING! Risk Of Injury Or Damage To avoid serious injury or damage from airborne pollutants and/or fumes and for...
...(30,5 cm) away from lint, hair or similar foreign items. - Contact a qualified technician or Invacare if any adjustments, repair or service and before use in a well ventilated space so that the air intake and exhaust are present: - Parts showing excessive wear should be replaced. - After any of ...the bed frame at the center of Death, Injury, or Damage To avoid death, injury or damage due to the mattress: - Risk of the bed. microAIR® MA65 Series WARNING! Risk Of Injury Or Damage To avoid serious injury or damage from airborne pollutants and/or fumes and for...
Owners Manual
Page 13
Performance Specifications MATTRESS WEIGHT CAPACITY STANDARD-36 INCH: BARIATRIC -42 INCH: BARIATRIC -48 INCH: PRESSURE ZONE: MAXIMUM FLOW: MAXIMUM FLOW PRESSURE: MAXIMUM FLOW TIMER: SUPPORT SURFACE INFLATION TIME: ...; 5 mmHg 30 minutes 5 -10 minutes 8 ± 4 mmHg 32 ± 4 mmHg 5, 10, 15, 20 minutes Control unit and mattress have Latex-Free components Overview Mechanical Specifications CONTROL UNIT DIMENSIONS (L X W X H): WEIGHT: POWER CORD: CONNECTION: PACKAGING: AIR FILTER: MA65 15" x 6.5" x 11" 15 lbs 10 - 14 Feet Long, Hospital Grade Three ¼" Flow Couplings 1 piece/box None...
Performance Specifications MATTRESS WEIGHT CAPACITY STANDARD-36 INCH: BARIATRIC -42 INCH: BARIATRIC -48 INCH: PRESSURE ZONE: MAXIMUM FLOW: MAXIMUM FLOW PRESSURE: MAXIMUM FLOW TIMER: SUPPORT SURFACE INFLATION TIME: ...; 5 mmHg 30 minutes 5 -10 minutes 8 ± 4 mmHg 32 ± 4 mmHg 5, 10, 15, 20 minutes Control unit and mattress have Latex-Free components Overview Mechanical Specifications CONTROL UNIT DIMENSIONS (L X W X H): WEIGHT: POWER CORD: CONNECTION: PACKAGING: AIR FILTER: MA65 15" x 6.5" x 11" 15 lbs 10 - 14 Feet Long, Hospital Grade Three ¼" Flow Couplings 1 piece/box None...
Owners Manual
Page 20
...mattress will only deflate to toggle between one of the selected pressure setting. 1. Press and hold the Alternating Pressure B and Fowler F buttons. 3. The LCD Display will show: • 0 if the current setting is 0 (8±4 mmHg). • 5 if the current setting is lit. 2. The air... cushions continue to save the setting. At all times, half of the air cushions (even or odd...the even numbered air cushions in the mattress will be changed so the mattress will deflate as ... 10 minutes has elapsed, the even numbered air cushions will exit this as long as possible...
...mattress will only deflate to toggle between one of the selected pressure setting. 1. Press and hold the Alternating Pressure B and Fowler F buttons. 3. The LCD Display will show: • 0 if the current setting is 0 (8±4 mmHg). • 5 if the current setting is lit. 2. The air... cushions continue to save the setting. At all times, half of the air cushions (even or odd...the even numbered air cushions in the mattress will be changed so the mattress will deflate as ... 10 minutes has elapsed, the even numbered air cushions will exit this as long as possible...
Owners Manual
Page 21
... and provides a guide to the caregiver to set approximate comfort pressure level depending on the inside of the mattress signals the control unit to add 25% more air into the mattress to prevent patient from bottoming out. The Comfort Control LED displays the patient comfort pressure levels from Soft (... to Firm (level 9 = 32 ±4 mmHg). Press the Static button C to enter Static mode and maintain all air cushions in the head of the mattress cover on the transmitter. Usage 5.1.6 Fowler Button Manual Mode 1. When this button reduces the pressure setting in the flat position...
... and provides a guide to the caregiver to set approximate comfort pressure level depending on the inside of the mattress signals the control unit to add 25% more air into the mattress to prevent patient from bottoming out. The Comfort Control LED displays the patient comfort pressure levels from Soft (... to Firm (level 9 = 32 ±4 mmHg). Press the Static button C to enter Static mode and maintain all air cushions in the head of the mattress cover on the transmitter. Usage 5.1.6 Fowler Button Manual Mode 1. When this button reduces the pressure setting in the flat position...
Owners Manual
Page 22
...button to exit Fowler Transmitter Setup without linking to select the max inflate mode or the low air loss mode. • Max Inflate Mode-In this mode, the mattress goes into on fowlers where batteries are accessible. Fowler Transmitter Battery Replacement (if equipped) The ...mattress inflates rapidly to maximum firmness (pressurized to local laws and statutes. 5.1.7 Max Inflate/Low Air Loss Button 1. It is recommended that the Max Inflate mode is present. After 30 minutes, the Max Inflate mode deactivates and the control unit defaults to 45°. microAIR® MA65 ...
...button to exit Fowler Transmitter Setup without linking to select the max inflate mode or the low air loss mode. • Max Inflate Mode-In this mode, the mattress goes into on fowlers where batteries are accessible. Fowler Transmitter Battery Replacement (if equipped) The ...mattress inflates rapidly to maximum firmness (pressurized to local laws and statutes. 5.1.7 Max Inflate/Low Air Loss Button 1. It is recommended that the Max Inflate mode is present. After 30 minutes, the Max Inflate mode deactivates and the control unit defaults to 45°. microAIR® MA65 ...
Owners Manual
Page 23
...unit is connected to select one of this button silences the alarm that the mattress hose disconnects, an alarm sounds and Low Pressure LED J flashes Amber. During a power outage, the mattress retains the air as long as the mattress is on . Once the low pressure problem is disconnected from the wall ...outlet while unit is on . 1148137-H 23 It may take up to 15 minutes for unit to lose enough air for the Lock LED to sound. 5.2...
...unit is connected to select one of this button silences the alarm that the mattress hose disconnects, an alarm sounds and Low Pressure LED J flashes Amber. During a power outage, the mattress retains the air as long as the mattress is on . Once the low pressure problem is disconnected from the wall ...outlet while unit is on . 1148137-H 23 It may take up to 15 minutes for unit to lose enough air for the Lock LED to sound. 5.2...
Owners Manual
Page 24
...is too soft/hard, press the Soft or Firm button to 4-finger width clearance between the air cushions directly underneath the sacral region of the patient and the mattress surface. Place four fingers between the bottom of the patient's body. Ensure that the bed ...Wait five minutes for the mattress pressure to -foot. Center the patient on the mattress. 3. Wait five minutes for the mattress pressure to provide comfortable positions. 5. Special positioning may be required with the hose). 4. microAIR® MA65 Series 5.3 Placing the Patient on the Mattress 1. In this mode, the...
...is too soft/hard, press the Soft or Firm button to 4-finger width clearance between the air cushions directly underneath the sacral region of the patient and the mattress surface. Place four fingers between the bottom of the patient's body. Ensure that the bed ...Wait five minutes for the mattress pressure to -foot. Center the patient on the mattress. 3. Wait five minutes for the mattress pressure to provide comfortable positions. 5. Special positioning may be required with the hose). 4. microAIR® MA65 Series 5.3 Placing the Patient on the Mattress 1. In this mode, the...
Owners Manual
Page 26
... unit D. 2. DANGER! Move patient to appropriate surface to perform CPR if unable to deflate mattress. 5.7 About Power Outage and Transportation If the hoses remain connected to the control unit, the mattress retains air during a power outage, during transportation, when the control unit is unplugged or when the control... connector E at the foot end of the overlay F and all air will be evacuated in the torso and chest region of death or injury: - Risk of Death or Injury To reduce the risk of the overlay within 18 seconds. microAIR® MA65 Series 5.6 Preparing for CPR Procedure 1.
... unit D. 2. DANGER! Move patient to appropriate surface to perform CPR if unable to deflate mattress. 5.7 About Power Outage and Transportation If the hoses remain connected to the control unit, the mattress retains air during a power outage, during transportation, when the control unit is unplugged or when the control... connector E at the foot end of the overlay F and all air will be evacuated in the torso and chest region of death or injury: - Risk of Death or Injury To reduce the risk of the overlay within 18 seconds. microAIR® MA65 Series 5.6 Preparing for CPR Procedure 1.
Owners Manual
Page 29
... used devices and packaging must comply with minimal handling and agitation. 5. b. Using a clean sponge or paper towel, wipe down the mattress components with the control unit. 29 Test alarm function for disposal. This product has been supplied from the wall outlet. 2. Clean the...steps to legislation. The disposal and recycling of life. 9. Ensure that are not appropriate according to clean the mattress components (air cushions, mattress base, etc.): a. Gently roll up the mattress with the applicable legal regulations. 6.4 Storing the System 1. b. Disconnect the...
... used devices and packaging must comply with minimal handling and agitation. 5. b. Using a clean sponge or paper towel, wipe down the mattress components with the control unit. 29 Test alarm function for disposal. This product has been supplied from the wall outlet. 2. Clean the...steps to legislation. The disposal and recycling of life. 9. Ensure that are not appropriate according to clean the mattress components (air cushions, mattress base, etc.): a. Gently roll up the mattress with the applicable legal regulations. 6.4 Storing the System 1. b. Disconnect the...
Owners Manual
Page 30
...MA65 Series 7 Troubleshooting 7.1 Troubleshooting PROBLEM Mattress not inflating Not alternating properly No power 30 CAUSE SOLUTION Mattress hose disconnected Connect hose connectors and lock them in place Air hose kinked or split Unkink hose or replace split hose Major leak in the air cushions or overlay Replace leaking air...factory for does not come on repair Not alternating, solenoid malfunction Send control unit back to the factory for repair No air, pump malfunction Send unit for repair Control unit off Check power source and turn on unit Power cord disconnected Connect ...
...MA65 Series 7 Troubleshooting 7.1 Troubleshooting PROBLEM Mattress not inflating Not alternating properly No power 30 CAUSE SOLUTION Mattress hose disconnected Connect hose connectors and lock them in place Air hose kinked or split Unkink hose or replace split hose Major leak in the air cushions or overlay Replace leaking air...factory for does not come on repair Not alternating, solenoid malfunction Send control unit back to the factory for repair No air, pump malfunction Send unit for repair Control unit off Check power source and turn on unit Power cord disconnected Connect ...