Owners Manual
Page 1
BEFORE using this product, read this manual and save for future reference. microAIR® MA90 and MA95 Series 90Z, 90ZB42, 90ZB48, 95Z, 95ZB42, 95ZB48 en Rotation with Alternating Pressure and Low Air Loss Mattress User Manual This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference. microAIR® MA90 and MA95 Series 90Z, 90ZB42, 90ZB48, 95Z, 95ZB42, 95ZB48 en Rotation with Alternating Pressure and Low Air Loss Mattress User Manual This manual MUST be given to the user of the product.
Owners Manual
Page 4
... damage. Caution indicates a potentially hazardous situation which could result in death or serious injury. Gives useful tips, recommendations and information of efficient, trouble-free use. microAIR® MA90 and MA95 Series 1 General 1.1 Symbols Signal words are used in this manual and apply to hazards or unsafe practices which , if not avoided...
... damage. Caution indicates a potentially hazardous situation which could result in death or serious injury. Gives useful tips, recommendations and information of efficient, trouble-free use. microAIR® MA90 and MA95 Series 1 General 1.1 Symbols Signal words are used in this manual and apply to hazards or unsafe practices which , if not avoided...
Owners Manual
Page 6
.... - Risk Of Injury Or Death To avoid injury or death from choking or ingestion of product: - DANGER! Always consult the patient's physician before using the MicroAir mattress system. 6 1148139-H DANGER! Risk Of Death Or Injury To avoid risk of death or injury from misuse of small parts or materials: - Risk Of... of death or injury due to misuse or modification: - Consult your therapist, physician and equipment supplier to pressure ulcers: - DO NOT use of the product. microAIR® MA90 and MA95 Series DANGER! Contraindications WARNING!
.... - Risk Of Injury Or Death To avoid injury or death from choking or ingestion of product: - DANGER! Always consult the patient's physician before using the MicroAir mattress system. 6 1148139-H DANGER! Risk Of Death Or Injury To avoid risk of death or injury from misuse of small parts or materials: - Risk Of... of death or injury due to misuse or modification: - Consult your therapist, physician and equipment supplier to pressure ulcers: - DO NOT use of the product. microAIR® MA90 and MA95 Series DANGER! Contraindications WARNING!
Owners Manual
Page 8
microAIR® MA90 and MA95 Series WARNING! Risk of Death, Injury or ... or damage from electrical components such as batteries and circuit boards. - WARNING! User manuals are available at www.invacare.com for repair if leakage is tightened securely. 8 1148139-H Entrapment DANGER! Variations in bed rail dimensions, and ...of entrapment. - Proper patient assessment and monitoring is required to the Bed Rail Entrapment Risk Notification Guide at www.invacare.com or your dealer. - Refer to reduce the risk of entrapment. - Work with therapist, physician and other...
microAIR® MA90 and MA95 Series WARNING! Risk of Death, Injury or ... or damage from electrical components such as batteries and circuit boards. - WARNING! User manuals are available at www.invacare.com for repair if leakage is tightened securely. 8 1148139-H Entrapment DANGER! Variations in bed rail dimensions, and ...of entrapment. - Proper patient assessment and monitoring is required to the Bed Rail Entrapment Risk Notification Guide at www.invacare.com or your dealer. - Refer to reduce the risk of entrapment. - Work with therapist, physician and other...
Owners Manual
Page 10
.... Keep openings free from airborne pollutants and/or fumes and for optimal performance: - DO NOT strap the mattress to collapse or missing hardware: - microAIR® MA90 and MA95 Series WARNING! Locate and position the product in well ventilated area. - Secure mattress straps to the bed deck at the ... foot ends and to the bed frame at least 12 inches (30,5 cm) away from walls, draperies and furniture. Contact a qualified technician or Invacare if any adjustments, repair or service and before use in this manual. WARNING! Keep product at the center of Death, Injury, or Damage To ...
.... Keep openings free from airborne pollutants and/or fumes and for optimal performance: - DO NOT strap the mattress to collapse or missing hardware: - microAIR® MA90 and MA95 Series WARNING! Locate and position the product in well ventilated area. - Secure mattress straps to the bed deck at the ... foot ends and to the bed frame at least 12 inches (30,5 cm) away from walls, draperies and furniture. Contact a qualified technician or Invacare if any adjustments, repair or service and before use in this manual. WARNING! Keep product at the center of Death, Injury, or Damage To ...
Owners Manual
Page 12
microAIR® MA90 and MA95 Series 3.2 Technical Data MA90Z power unit is also used on MA90ZB42 and MA90ZB48 systems. MA95Z power unit is also used on MA95ZB42 and MA95ZB48 systems. Electrical Specifications Input Voltage AC: Input Frequency: Current: Maximum Power Consumption: Circuit Protection: Mode Of Operation: MA90Z MA95Z... Temperature: Relative Humidity: Atmospheric Pressure: Storage And Shipping Conditions Ambient Temperature: Relative Humidity: Atmospheric Pressure: MA90Z/MA95Z 50°-95° F 30%-75% Non-Condensing 70-106 kPa -40°-158° F 10%-100% 50-106 kPa 12...
microAIR® MA90 and MA95 Series 3.2 Technical Data MA90Z power unit is also used on MA90ZB42 and MA90ZB48 systems. MA95Z power unit is also used on MA95ZB42 and MA95ZB48 systems. Electrical Specifications Input Voltage AC: Input Frequency: Current: Maximum Power Consumption: Circuit Protection: Mode Of Operation: MA90Z MA95Z... Temperature: Relative Humidity: Atmospheric Pressure: Storage And Shipping Conditions Ambient Temperature: Relative Humidity: Atmospheric Pressure: MA90Z/MA95Z 50°-95° F 30%-75% Non-Condensing 70-106 kPa -40°-158° F 10%-100% 50-106 kPa 12...
Owners Manual
Page 14
microAIR® MA90 and MA95 Series Mechanical Specifications Control Unit Dimensions (L X W X H): Control Unit Weight: Control Unit Power Cord: Control Unit Connection: Control Unit Packaging: Air Filter: MA90Z MA95Z 15" x 6½" x 11" 15 lbs 10-14 Feet Long, Hospital Grade Quad Flow-Four ¼" Single Quick Coupling Triple Flow-Two ½" and One ¼" Single Magnetic Quick Coupling 1 piece/box None Charcoal Air Filter with Fire Retardant 14 1148139-H
microAIR® MA90 and MA95 Series Mechanical Specifications Control Unit Dimensions (L X W X H): Control Unit Weight: Control Unit Power Cord: Control Unit Connection: Control Unit Packaging: Air Filter: MA90Z MA95Z 15" x 6½" x 11" 15 lbs 10-14 Feet Long, Hospital Grade Quad Flow-Four ¼" Single Quick Coupling Triple Flow-Two ½" and One ¼" Single Magnetic Quick Coupling 1 piece/box None Charcoal Air Filter with Fire Retardant 14 1148139-H
Owners Manual
Page 16
... it on the bed frame. Remove the original foam mattress from the bed. 2. Ensure that the hose E is towards the foot end B of the bed. microAIR® MA90 and MA95 Series 4.2 Mattress Replacement System Installation CAUTION!
... it on the bed frame. Remove the original foam mattress from the bed. 2. Ensure that the hose E is towards the foot end B of the bed. microAIR® MA90 and MA95 Series 4.2 Mattress Replacement System Installation CAUTION!
Owners Manual
Page 18
Perform one of the mattress. 3. The audible click indicates that hose connector(s) are properly engaged with control connectors. 18 1148139-H Locate the hose at the foot end of the following: • MA90Z Power Unit-Perform the following: 2. a. Squeeze and hold the tabs C on right side of control unit B. Push the hose connectors onto the control connector until an audible click is heard. Locate control unit connectors on the hose connector D. microAIR® MA90 and MA95 Series 1. b.
Perform one of the mattress. 3. The audible click indicates that hose connector(s) are properly engaged with control connectors. 18 1148139-H Locate the hose at the foot end of the following: • MA90Z Power Unit-Perform the following: 2. a. Squeeze and hold the tabs C on right side of control unit B. Push the hose connectors onto the control connector until an audible click is heard. Locate control unit connectors on the hose connector D. microAIR® MA90 and MA95 Series 1. b.
Owners Manual
Page 20
... causes erratic behavior, unplug the product immediately. Other devices may be reduced or corrected by the above standard. Contact Invacare for any additional EMC information or EMC tables for Type BF equipment. Leave unplugged while transmission is causing the interference....protection against electromagnetic interference in a professional healthcare environment. To determine if the emissions from even the low levels of this manual. microAIR® MA90 and MA95 Series 4.7 EMC Information Use of the following measures: • Reposition, relocate, or increase the separation ...
... causes erratic behavior, unplug the product immediately. Other devices may be reduced or corrected by the above standard. Contact Invacare for any additional EMC information or EMC tables for Type BF equipment. Leave unplugged while transmission is causing the interference....protection against electromagnetic interference in a professional healthcare environment. To determine if the emissions from even the low levels of this manual. microAIR® MA90 and MA95 Series 4.7 EMC Information Use of the following measures: • Reposition, relocate, or increase the separation ...
Owners Manual
Page 22
... rotation modes or to set the unit to select one of the air cushions are maintained at a constant pressure. 5.1.6 Function Button (MA95Z Power Unit) 1. Press the Function button F to move the pointer on the LCD to select Static mode or Alternating Rotation mode. ...are : 22 1148139-H Press the Rotation button E to Function. 2. The rotation times are maintained at a constant pressure. 5.1.5 Rotation Button (MA95Z Power Unit) 1. microAIR® MA90 and MA95 Series 5.1.1 Power Button 1. Press the Firm or Soft button until the desired function displays. Once the Power button is...
... rotation modes or to set the unit to select one of the air cushions are maintained at a constant pressure. 5.1.6 Function Button (MA95Z Power Unit) 1. Press the Function button F to move the pointer on the LCD to select Static mode or Alternating Rotation mode. ...are : 22 1148139-H Press the Rotation button E to Function. 2. The rotation times are maintained at a constant pressure. 5.1.5 Rotation Button (MA95Z Power Unit) 1. microAIR® MA90 and MA95 Series 5.1.1 Power Button 1. Press the Firm or Soft button until the desired function displays. Once the Power button is...
Owners Manual
Page 24
microAIR® MA90 and MA95 Series 3. Use the Firmer or Softer buttons to height ratio is ...Inflate mode deactivates and the control unit defaults to the previous setting. • It takes 25-60 seconds for the MA95Z mattress to set approximate comfort pressure level depending on -demand low air loss relief mode. A series of mattress). 1148139...can be used during patient ingress/egress, patient wound care, patient turning or patient cleaning. 5.1.10 Max Inflate Button (MA95Z Power Unit) Press the Max Inflate button K to select the max inflate mode and inflate the mattress rapidly to maximum...
microAIR® MA90 and MA95 Series 3. Use the Firmer or Softer buttons to height ratio is ...Inflate mode deactivates and the control unit defaults to the previous setting. • It takes 25-60 seconds for the MA95Z mattress to set approximate comfort pressure level depending on -demand low air loss relief mode. A series of mattress). 1148139...can be used during patient ingress/egress, patient wound care, patient turning or patient cleaning. 5.1.10 Max Inflate Button (MA95Z Power Unit) Press the Max Inflate button K to select the max inflate mode and inflate the mattress rapidly to maximum...
Owners Manual
Page 26
... power outage. Replace fowler if red indicator is first inserted the LED light will need to local laws and statutes. 5.1.12 Lock/Alarm Silence Button 1. microAIR® MA90 and MA95 Series Fowler Transmitter Battery Replacement (if equipped) The transmitter has an LED light to prevent any tampering with the settings. Green...
... power outage. Replace fowler if red indicator is first inserted the LED light will need to local laws and statutes. 5.1.12 Lock/Alarm Silence Button 1. microAIR® MA90 and MA95 Series Fowler Transmitter Battery Replacement (if equipped) The transmitter has an LED light to prevent any tampering with the settings. Green...
Owners Manual
Page 28
... be required with the hose). 4. Ensure that the bed is achieved. 10. b. Ensure that the patient's feet are towards the foot end of the mattress. d. microAIR® MA90 and MA95 Series 5.3 Placing Patient on the bed. 3. Adjust the comfort setting, if needed.
... be required with the hose). 4. Ensure that the bed is achieved. 10. b. Ensure that the patient's feet are towards the foot end of the mattress. d. microAIR® MA90 and MA95 Series 5.3 Placing Patient on the bed. 3. Adjust the comfort setting, if needed.
Owners Manual
Page 30
Pull the hose connector B from the control unit connectors. 3. Disconnect the red CPR connector located on the hose connector B . 2. microAIR® MA90 and MA95 Series 5.6 Preparing for CPR Procedure MA90Z 1. Press and hold the tabs A on the hose F. 30 1148139-H
Pull the hose connector B from the control unit connectors. 3. Disconnect the red CPR connector located on the hose connector B . 2. microAIR® MA90 and MA95 Series 5.6 Preparing for CPR Procedure MA90Z 1. Press and hold the tabs A on the hose F. 30 1148139-H
Owners Manual
Page 32
... 1148139-H Ensure that the control unit is occupying the bed. - Before cleaning or disassembling the MA90 and MA95Z series, check the underside of all equipment. Do not clean the bed when the patient is off. 4. microAIR® MA90 and MA95 Series 6 Maintenance 6.1 Service Life WARNING! This product has an expected lifetime of...
... 1148139-H Ensure that the control unit is occupying the bed. - Before cleaning or disassembling the MA90 and MA95Z series, check the underside of all equipment. Do not clean the bed when the patient is off. 4. microAIR® MA90 and MA95 Series 6 Maintenance 6.1 Service Life WARNING! This product has an expected lifetime of...
Owners Manual
Page 34
... of in operation. Disconnect the air hose connector from the control unit and allow air to clean the mattress components (air cushions, mattress base, etc.): a. microAIR® MA90 and MA95 Series 9. b. Air dry all crevices as they can harbor microorganisms. c. This product may contain substances that complies with the individual product...
... of in operation. Disconnect the air hose connector from the control unit and allow air to clean the mattress components (air cushions, mattress base, etc.): a. microAIR® MA90 and MA95 Series 9. b. Air dry all crevices as they can harbor microorganisms. c. This product may contain substances that complies with the individual product...
Owners Manual
Page 36
... power source and turn it On Power outage Wait till the power source has power Blown fuse Replace blown fuse with an equivalent fuse 1148139-H microAIR® MA90 and MA95 Series 7 Troubleshooting 7.1 Troubleshooting Problem Mattress not inflating Not alternating properly No power 36 Cause Solution Mattress hose disconnected Connect hose connectors...
... power source and turn it On Power outage Wait till the power source has power Blown fuse Replace blown fuse with an equivalent fuse 1148139-H microAIR® MA90 and MA95 Series 7 Troubleshooting 7.1 Troubleshooting Problem Mattress not inflating Not alternating properly No power 36 Cause Solution Mattress hose disconnected Connect hose connectors...
Owners Manual
Page 38
...a copy of title to Invacare at Invacare option. Invacare will be repaired or replaced, at the address on the back cover. IN ADDITION, THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED; microAIR® MA90 and MA95 Series... serial number. The internal pump, blower and compressor are warranted for a year from the date of purchase from Invacare or a dealer. Invacare's sole obligation and your Invacare product. Provide dealer's name, address, model number, and the date of purchase, indicate nature of return authorization...
...a copy of title to Invacare at Invacare option. Invacare will be repaired or replaced, at the address on the back cover. IN ADDITION, THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED; microAIR® MA90 and MA95 Series... serial number. The internal pump, blower and compressor are warranted for a year from the date of purchase from Invacare or a dealer. Invacare's sole obligation and your Invacare product. Provide dealer's name, address, model number, and the date of purchase, indicate nature of return authorization...