Fluke 355 Support Question
Find answers below for this question about Fluke 355.Need a Fluke 355 manual? We have 2 online manuals for this item!
Question posted by bennykittoop on August 13th, 2015
Clamp Meter Out Of Accuracy
fluke model 355 out of accuracy showing more
Current Answers
Answer #1: Posted by waelsaidani1 on September 6th, 2015 10:38 AM
troubleshooting variable speed drives (VSD), the input bandwidth of a meter can become very important toward getting a meaningful reading. Due to the high harmonic content in the signal coming out of a VSD to the motor, a DMM would measure most of the voltage content depending on its input bandwidth. Measuring the voltage output of a VSD is not a common measurement; if you did the reading would be higher. A motor connected to a VSD only responds to the average value of the signal, and to measure that power, the input bandwidth of the meter must be narrower than its DMM counterpart. The Fluke 337 Clamp Meter has been specifically designed for testing and troubleshooting VSDs.
How to make voltage measurements
- Select Volts AC (V~) or Volts DC (V ), as desired.
- Plug the black test probe into the COM input jack. Plug the red test probe into the V input jack.
- Touch the probe tips to the circuit across a load or power source (in parallel to the circuit).
- View the reading, being sure to note the unit of measurement.
- (Optional) Press the HOLD button to freeze the reading in the display. Now you can remove the meter from the live circuit and then read the display when you are safely clear of the electrical hazard.
Related Fluke 355 Manual Pages
Similar Questions
Cal Err Message Is Showing While On The Clamp Meter
CAL Err message is showing while power on the clamp meter.No buttons are functioning after this erro...
CAL Err message is showing while power on the clamp meter.No buttons are functioning after this erro...
(Posted by kvramesh 1 year ago)
When I Measure The Voltage, The Voltage Is Not Showing Why?
(Posted by yogendrashetty12 3 years ago)
Pantalla Remota Flete 381
Buenos días donde puedo conseguir la pantalla inalámbrica para la pinza amperica 381
Buenos días donde puedo conseguir la pantalla inalámbrica para la pinza amperica 381
(Posted by delossantos2577 3 years ago)
Calibration Adjustment Procedure For Fluke 355 Clamp Meter
clamp meter readings out specification we need to know how to minimize the error
clamp meter readings out specification we need to know how to minimize the error
(Posted by bennykittoop 8 years ago)