27106T Gill Propeller Anemometer
Page 2
...handle repairs for our employees. To obtain a Returned Materials Authorization (RMA), contact Campbell Scientific, Inc., phone (435) 753-2342. Campbell Scientific, Inc. products which Campbell Scientific company serves your country. This warranty is in it. The following contact information is... number will return such products by Campbell Scientific, Inc. Campbell Scientific will be issued. directly. Please write this number clearly on products that were exposed to Campbell Scientific, Inc. Warranty and Assistance The 27106T GILL PROPELLER ANEMOMETER is warranted by ...
...handle repairs for our employees. To obtain a Returned Materials Authorization (RMA), contact Campbell Scientific, Inc., phone (435) 753-2342. Campbell Scientific, Inc. products which Campbell Scientific company serves your country. This warranty is in it. The following contact information is... number will return such products by Campbell Scientific, Inc. Campbell Scientific will be issued. directly. Please write this number clearly on products that were exposed to Campbell Scientific, Inc. Warranty and Assistance The 27106T GILL PROPELLER ANEMOMETER is warranted by ...
27106T Gill Propeller Anemometer
Page 3
... Contents PDF viewers note: These page numbers refer to Campbell Scientific Dataloggers 5 5-1. Wiring 5 5. Sensor Maintenance 8 6.1 Flange Bearing Replacement 8 6.2 Tach-Generator Replacement 9 7. Wind Speed Multiplier 6 5-2. Connections to the printed version of this document. Wiring for links to Crossarm 3 4. Installation 2 3.1 Unpack Equipment and Assemble the 27106T 2 3.2 Mounting to specific sections. 1. Troubleshooting 10 8. General Description 1 2. Programming...
... Contents PDF viewers note: These page numbers refer to Campbell Scientific Dataloggers 5 5-1. Wiring 5 5. Sensor Maintenance 8 6.1 Flange Bearing Replacement 8 6.2 Tach-Generator Replacement 9 7. Wind Speed Multiplier 6 5-2. Connections to the printed version of this document. Wiring for links to Crossarm 3 4. Installation 2 3.1 Unpack Equipment and Assemble the 27106T 2 3.2 Mounting to specific sections. 1. Troubleshooting 10 8. General Description 1 2. Programming...
27106T Gill Propeller Anemometer
Page 6
...) and cabled by Campbell Scientific for axial flows. The shipping carton should include: • 16" x 2" x 2" box labeled Propeller Anemometer • 9" x 9" x 2" box labeled Carbon Fiber Propeller • Cable (routed through the 1180 mounting pipe to mating connector) • 1180 Mounting Pipe CAUTION Make sure that you have all of the 27106T's components. This compound...
...) and cabled by Campbell Scientific for axial flows. The shipping carton should include: • 16" x 2" x 2" box labeled Propeller Anemometer • 9" x 9" x 2" box labeled Carbon Fiber Propeller • Cable (routed through the 1180 mounting pipe to mating connector) • 1180 Mounting Pipe CAUTION Make sure that you have all of the 27106T's components. This compound...
27106T Gill Propeller Anemometer
Page 9
... this sensor can be wired to obtain miles/hour, meters/second, km/H, and knots. 4. Wiring 27106T Gill Propeller Anemometer Connections to Campbell Scientific Dataloggers Color White Black Clear Wire Label CR800 CR850 CR5000 CR3000 CR1000 Signal SE Analog Signal Reference Shield CR510... is used to create the datalogger program, the sensor should be generated using Campbell Scientific's Short Cut Program Builder software. When Short Cut for wind speed (U) is zero. 5 Connections to Campbell Scientific dataloggers are given in the wiring diagram created by the Volt (SE) (...
... this sensor can be wired to obtain miles/hour, meters/second, km/H, and knots. 4. Wiring 27106T Gill Propeller Anemometer Connections to Campbell Scientific Dataloggers Color White Black Clear Wire Label CR800 CR850 CR5000 CR3000 CR1000 Signal SE Analog Signal Reference Shield CR510... is used to create the datalogger program, the sensor should be generated using Campbell Scientific's Short Cut Program Builder software. When Short Cut for wind speed (U) is zero. 5 Connections to Campbell Scientific dataloggers are given in the wiring diagram created by the Volt (SE) (...
27106T Gill Propeller Anemometer
Page 12
... to the factory. REMOVE OLD BEARINGS a) Remove propeller from generator housing. 8 Components are not available, return the instrument to Campbell Scientific and have their qualified technicians replace the bearing. b) Unthread and separate shaft housing assembly from anemometer. Replace bearings as follows: ... threshold increases above an acceptable level, bearings may need replacement due to be done by qualified technicians. 27106T Gill Propeller Anemometer 6: Minimum (P74) 1: 1 Reps 2: 0 Value Only 3: 1 Loc [ WS_ms ] 7: Average (P71)^4771 1: 1 Reps 2: ...
... to the factory. REMOVE OLD BEARINGS a) Remove propeller from generator housing. 8 Components are not available, return the instrument to Campbell Scientific and have their qualified technicians replace the bearing. b) Unthread and separate shaft housing assembly from anemometer. Replace bearings as follows: ... threshold increases above an acceptable level, bearings may need replacement due to be done by qualified technicians. 27106T Gill Propeller Anemometer 6: Minimum (P74) 1: 1 Reps 2: 0 Value Only 3: 1 Loc [ WS_ms ] 7: Average (P71)^4771 1: 1 Reps 2: ...
27106T Gill Propeller Anemometer
Page 13
... to confirm it is indeed caused by a qualified technician. When the tach-generator output becomes erratic (usually due to brush failure) or begins to Campbell Scientific and have their qualified technicians replace the TachGenerator. c) Carefully slide propeller shaft through both bearings and out of housing. e) Place shaft collar/coupling disk...end play gap between shaft collar/coupling disk and bearing. If replacing the tach-generator due to excessive torque make certain it into housing. 27106T Gill Propeller Anemometer c) Loosen set screw (80 oz in, 5600 gm-cm max torque).
... to confirm it is indeed caused by a qualified technician. When the tach-generator output becomes erratic (usually due to brush failure) or begins to Campbell Scientific and have their qualified technicians replace the TachGenerator. c) Carefully slide propeller shaft through both bearings and out of housing. e) Place shaft collar/coupling disk...end play gap between shaft collar/coupling disk and bearing. If replacing the tach-generator due to excessive torque make certain it into housing. 27106T Gill Propeller Anemometer c) Loosen set screw (80 oz in, 5600 gm-cm max torque).