BEFORE using this product, read this manual and save for future reference. Invacare® ML4, ML6 and M9 Cylinder with Conserver
For HomeFill® Oxygen System en HF2PCL4, HF2PCL6, HF2PC9
User Manual
This manual MUST be given to the user of the product.
Owners Manual - Page 3
...1.3 User Profile 5 1.4 Service Life 5 1.5 Limited Warranty 6 1.6 Component Identification 8
2 Safety 9 2.1 Label Locations 9 2.2 General Guidelines 10
3 Setup 15 3.1 Dealer Setup 15 3.2 User Setup 15 3.3 Receiving Inspection 15...22 6.1 Disposal 22 6.2 Wear and Tear 22
7 Troubleshooting 23 7.1 Troubleshooting 23
8 Technical Data 24 8.1 Technical Description 24 8.2 Specifications 24
Owners Manual - Page 4
...will result in personal injury or property damage.
Invacare® ML4, ML6 and M9 Cylinder with Conserver
1 General
1.1 Symbols
Signal words are used in this manual and apply to property if it is not ...in death or serious injury.
WARNING! - See the information below for efficient, trouble-free use.
4
Read the manual No open flame DO NOT Smoke Oxidizer Recycle DO NOT dispose of the signal...
Owners Manual - Page 5
... product is to be distributed to inhale and exhale without the use and specifications has not been tested and may increase the drying out of the product....settings and may lead to be used by patients who breathe through their mouths.
1.4 Service Life
WARNING!
WARNING! Only personnel instructed and trained in this manual. DANGER! This product is not intended to be life supporting...
Owners Manual - Page 6
...warranty.
Do not return products to the product instructions, .
This warranty only applies to the labor for repairs performed by Invacare in a manner to avoid shipping damage and to ship the unit to either Invacare's plant or service center as filters, are excluded from defects in replacement part installation or repair of any and all warranties.
otherwise, this warranty...
Owners Manual - Page 10
... of Death, Injury or Damage - Invacare® ML4, ML6 and M9 Cylinder with Conserver
2.2 General Guidelines
In order to ensure the safe installation, assembly and operation of the product these instructions and any additional instructional material such as user manuals, service manuals or instruction sheets supplied with the HomeFill
compressor user manual. - Check ALL external components and...
Owners Manual - Page 14
Refer to the HomeFill User Manual for additional warnings, cautions and safety instructions.
14
1112648-H-00 DO NOT use and by courier or postal service. ALWAYS confirm prescribed dose before administering to patient and monitor on outlet connection prior to use. This product is not in use a humidifier with a relief valve. Invacare® ML4, ML6...
Owners Manual - Page 15
...the prescribed level of cylinder assemblies is very important to be used with the HomeFill user manual. - Keep out of the reach of Injury or Damage - WARNING! Secure the ...there is to the Receiving Inspection section.
1112648-H-00
Setup
3.3 Receiving Inspection
Remove product from package and inspect for personal
use by Invacare HomeFill Oxygen System only. - Risk of children.
15...
Owners Manual - Page 17
...to refill the cylinder assembly on the Invacare HomeFill Compressor. 17 If for visible ...manual for service or replacement. 4 Usage
4.1 Before Use
WARNING! If below 1500 psig (10342 kPa), then refill the cylinder assembly using the HomeFill system.
All cylinders MUST be defective or does not
work properly, contact an Invacare dealer or qualified technician for operating instructions...
Owners Manual - Page 18
...areas of injury: - aisles, doorways, hallways, etc.)
DO NOT use the Invacare Corporation carrying bag.
10. Turn the flow dial E to
setting "Off".
18
1112648-H-00 Ensure there are no kinks or twists in your...
4.
Risk of Injury To reduce the risk of high foot traffic (i.e.
Risk of oxygen as instructed by your nose as you inhale. 11. The cannula MUST be routed and secured
properly. -...
Owners Manual - Page 19
...damage may occur. The conserver should be replaced at a minimum every 5 years. This inspection and test MUST be performed ONLY by an Invacare approved repair center. NEVER use near any person, or.... DO NOT use cleaning/disinfecting solutions on product. - Contact your provider for service or repair. - DO NOT use or fill cylinder assembly if the cylinder
fillport is necessary...
Owners Manual - Page 20
Invacare® ML4, ML6 and M9 Cylinder with 70% isopropyl alcohol. DO NOT completely empty the cylinder. Turn the flow dial to setting... flow dial E to setting "Off". 3. Follow the cleaning/disinfectant product manufacturer's instructions for disinfecting the cylinder ...cylinder assembly A so the outlet connection D is installed on the fillport. 2. Recommended disinfectant products:
...
Owners Manual - Page 21
...is dry before using.
5.5 Reuse
WARNING! Invacare products are specifically designed and
manufactured for use with Invacare accessories. To obtain Invacare accessories, contact Invacare by other sources of non-Invacare accessories may result in serious injury or damage. - The following procedures: - Before transferring the product to setting "Off". 2. Risk of Injury or Damage...
Owners Manual - Page 23
... "Off". 2. Fill or replace cylinder. 3. If problem cannot be corrected, consult your Dealer/Home Care equipment provider. 7 Troubleshooting
Troubleshooting
7.1 Troubleshooting
If the Regulating Cylinder fails to prescribed setting. 2. Conserver not sensing breath. 5.
Kinked or blocked cannula. Adjust flow dial to function, consult the Troubleshooting guide. Flow dial is kinked...
Owners Manual - Page 26
Invacare® ML4, ML6 and M9 Cylinder with Conserver
HF2PCL6 HF2PC9 *With Concentrator flow rates to patient: Concentrator Model Platinum 5 and... Model Single Lumen Cannula Carry Bag
HF2PCL4
HF2PCL6
HF2PC9
4' (1.21 m) recommended. 7' (2.1 m) tubing length maximum
HF2PCL4 Carry Bag
HF2PCL6 Carry Bag
HF2PC9 Carry Bag
Regulatory Compliance CGA G-4.1 (Cleaning Equipment For Oxygen Service)...