BEFORE using this product, read this manual and save for future reference. Invacare® ML4, ML6 and M9 Cylinder with Conserver For HomeFill® Oxygen System en HF2PCL4, HF2PCL6, HF2PC9 User Manual This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference. Invacare® ML4, ML6 and M9 Cylinder with Conserver For HomeFill® Oxygen System en HF2PCL4, HF2PCL6, HF2PC9 User Manual This manual MUST be given to the user of the product.
Owners Manual
Page 2
WD-40 and 3-IN-ONE are identified by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. All trademarks are owned by ™and ®. ©2016 Invacare®Corporation All rights reserved. Trademarks are registered trademarks of WD-40 Company.
WD-40 and 3-IN-ONE are identified by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. All trademarks are owned by ™and ®. ©2016 Invacare®Corporation All rights reserved. Trademarks are registered trademarks of WD-40 Company.
Owners Manual
Page 3
...Use 5 1.3 User Profile 5 1.4 Service Life 5 1.5 Limited Warranty 6 1.6 Component Identification 8 2 Safety 9 2.1 Label Locations 9 2.2 General Guidelines 10 3 Setup 15 3.1 Dealer Setup 15 3.2 User Setup 15 3.3 Receiving Inspection 15 3.4 Cylinder Prefill Inspection 15 3.4.1 The Hydrostatic Testing Date 16 3.4.2 External Examination 16 4 Usage 17 4.1 Before Use... 5.6 Storage 21 5.7 Optional Accessories 21 6 After Use 22 6.1 Disposal 22 6.2 Wear and Tear 22 7 Troubleshooting 23 7.1 Troubleshooting 23 8 Technical Data 24 8.1 Technical Description 24 8.2 Specifications 24
...Use 5 1.3 User Profile 5 1.4 Service Life 5 1.5 Limited Warranty 6 1.6 Component Identification 8 2 Safety 9 2.1 Label Locations 9 2.2 General Guidelines 10 3 Setup 15 3.1 Dealer Setup 15 3.2 User Setup 15 3.3 Receiving Inspection 15 3.4 Cylinder Prefill Inspection 15 3.4.1 The Hydrostatic Testing Date 16 3.4.2 External Examination 16 4 Usage 17 4.1 Before Use... 5.6 Storage 21 5.7 Optional Accessories 21 6 After Use 22 6.1 Disposal 22 6.2 Wear and Tear 22 7 Troubleshooting 23 7.1 Troubleshooting 23 8 Technical Data 24 8.1 Technical Description 24 8.2 Specifications 24
Owners Manual
Page 4
...or both. - WARNING! - Indicates a hazardous situation that could result in death or serious injury. Invacare® ML4, ML6 and M9 Cylinder with Conserver 1 General 1.1 Symbols Signal words are used in this manual and apply to property if it is not avoided. Caution indicates a potentially hazardous situation which , ... hazardous situation which , if not avoided, may result in personal injury or property damage. See the information below for efficient, trouble-free use. 4 Read the manual No open flame DO NOT Smoke Oxidizer Recycle DO NOT dispose of the signal words. CAUTION! -
...or both. - WARNING! - Indicates a hazardous situation that could result in death or serious injury. Invacare® ML4, ML6 and M9 Cylinder with Conserver 1 General 1.1 Symbols Signal words are used in this manual and apply to property if it is not avoided. Caution indicates a potentially hazardous situation which , ... hazardous situation which , if not avoided, may result in personal injury or property damage. See the information below for efficient, trouble-free use. 4 Read the manual No open flame DO NOT Smoke Oxidizer Recycle DO NOT dispose of the signal words. CAUTION! -
Owners Manual
Page 5
... safety instructions, maintenance intervals and correct use of this manual. 1.3 User Profile No specific product knowledge or training is required, other than what is contained in this user manual. ONLY use and specifications has not been tested and may increase the drying out of the product. - Perform all maintenance according to the recommended schedule in this product. The expected service...
... safety instructions, maintenance intervals and correct use of this manual. 1.3 User Profile No specific product knowledge or training is required, other than what is contained in this user manual. ONLY use and specifications has not been tested and may increase the drying out of the product. - Perform all maintenance according to the recommended schedule in this product. The expected service...
Owners Manual
Page 6
... user or shipping charges incurred in replacement part installation or repair of an alleged defect in a manner to avoid shipping damage and to ship the unit to such repair or replacement. Invacare's sole obligation and your exclusive remedy under this will void any such product shall be repaired or replaced, at Invacare's option. Do not return products to the product instructions, . For Model...
... user or shipping charges incurred in replacement part installation or repair of an alleged defect in a manner to avoid shipping damage and to ship the unit to such repair or replacement. Invacare's sole obligation and your exclusive remedy under this will void any such product shall be repaired or replaced, at Invacare's option. Do not return products to the product instructions, . For Model...
Owners Manual
Page 10
...Invacare® ML4, ML6 and M9 Cylinder with Conserver 2.2 General Guidelines In order to ensure the safe installation, assembly and operation of this product may cause death, injury or damage. Risk of Death, Injury or Damage Improper use this product or any additional instructional material such as user manuals, service manuals or instruction... the valve after each use the contents until you are not to use by the Invacare HomeFill oxygen system shall be followed. Risk of Death, Injury or Damage - Use this product. - Training required to be filled for repair. - This product is...
...Invacare® ML4, ML6 and M9 Cylinder with Conserver 2.2 General Guidelines In order to ensure the safe installation, assembly and operation of this product may cause death, injury or damage. Risk of Death, Injury or Damage Improper use this product or any additional instructional material such as user manuals, service manuals or instruction... the valve after each use the contents until you are not to use by the Invacare HomeFill oxygen system shall be followed. Risk of Death, Injury or Damage - Use this product. - Training required to be filled for repair. - This product is...
Owners Manual
Page 11
... difficulty is being delivered. - DO NOT clean with such substances, contact your provider for service or repair. - DO NOT use if dirt or contaminants are present on the order of products Invacare suggests you use these types of a medical practitioner licensed by Invacare. - Avoid creation of oxygen and cause injury. Keep all matches, lighted cigarettes, electronic...
... difficulty is being delivered. - DO NOT clean with such substances, contact your provider for service or repair. - DO NOT use if dirt or contaminants are present on the order of products Invacare suggests you use these types of a medical practitioner licensed by Invacare. - Avoid creation of oxygen and cause injury. Keep all matches, lighted cigarettes, electronic...
Owners Manual
Page 12
...turn, do not force, otherwise damage or injury could occur. - DO NOT use Invacare Corporation carrying bag. - WARNING! Risk of Injury or Damage To reduce the risk of the conserver. - Only use if the oxygen has an unusual odor. - DO NOT allow cylinder assemblies to...be performed by Invacare Corporation or in an Invacare approved repair center. - Risk of liquid. - DO NOT immerse product in upright position when not connected to impact inanimate objects. 1112648-H-00 All repairs MUST be turned with Conserver DANGER! DO NOT use a flow outlet for service. Risk of ...
...turn, do not force, otherwise damage or injury could occur. - DO NOT use Invacare Corporation carrying bag. - WARNING! Risk of Injury or Damage To reduce the risk of the conserver. - Only use if the oxygen has an unusual odor. - DO NOT allow cylinder assemblies to...be performed by Invacare Corporation or in an Invacare approved repair center. - Risk of liquid. - DO NOT immerse product in upright position when not connected to impact inanimate objects. 1112648-H-00 All repairs MUST be turned with Conserver DANGER! DO NOT use a flow outlet for service. Risk of ...
Owners Manual
Page 14
Invacare® ML4, ML6 and M9 Cylinder with Conserver The cylinder assembly MUST be emptied of product, turn the flow dial to the "Off" setting, and contact your dealer/Home Care equipment provider. Be sure all connections are indicated by loss of 2.1 m (7 ft). NO OXYGEN is delivered when the conserver flow dial setting... is not suitable to be installed on a frequent basis. The cannula is for additional warnings, cautions and safety instructions...., discontinue use of gas prior to the HomeFill User Manual for single patient use and by courier or postal service. ALWAYS...
Invacare® ML4, ML6 and M9 Cylinder with Conserver The cylinder assembly MUST be emptied of product, turn the flow dial to the "Off" setting, and contact your dealer/Home Care equipment provider. Be sure all connections are indicated by loss of 2.1 m (7 ft). NO OXYGEN is delivered when the conserver flow dial setting... is not suitable to be installed on a frequent basis. The cannula is for additional warnings, cautions and safety instructions...., discontinue use of gas prior to the HomeFill User Manual for single patient use and by courier or postal service. ALWAYS...
Owners Manual
Page 15
...Before the first patient use , the provider MUST check the hydrostatic testing date. Before patient use , the provider MUST disinfect the product. Refer to the Disinfecting section. - Refer to be used with the HomeFill user manual. - DO NOT drop...using the pressure swing adsorption process. - 3 Setup 3.1 Dealer Setup WARNING! - Uninterrupted use by professional users. - This cylinder contains 93±3% oxygen produced by Invacare HomeFill Oxygen System only. - Secure the cylinder assembly during storage and use . Contact your physician or therapist. 3.2 User Setup...
...Before the first patient use , the provider MUST check the hydrostatic testing date. Before patient use , the provider MUST disinfect the product. Refer to the Disinfecting section. - Refer to be used with the HomeFill user manual. - DO NOT drop...using the pressure swing adsorption process. - 3 Setup 3.1 Dealer Setup WARNING! - Uninterrupted use by professional users. - This cylinder contains 93±3% oxygen produced by Invacare HomeFill Oxygen System only. - Secure the cylinder assembly during storage and use . Contact your physician or therapist. 3.2 User Setup...
Owners Manual
Page 16
...replacement. 3. DO NOT use caustic paint strippers. DANGER! Valve and safety relief devices MUST be withdrawn from Fire To avoid fire, death, injury or damage: - otherwise, injury or damage may result. Invacare...replaced by trained personnel. DO NOT use oil or grease. DO NOT alter or change valve assembly. - Risk of Injury To reduce the risk of oil and grease to follow these safety instructions..., Injury or Damage from service and condemned by trained and authorized personnel. Contact your dealer/Home Care equipment provider for service. • Dents or ...
...replacement. 3. DO NOT use caustic paint strippers. DANGER! Valve and safety relief devices MUST be withdrawn from Fire To avoid fire, death, injury or damage: - otherwise, injury or damage may result. Invacare...replaced by trained personnel. DO NOT use oil or grease. DO NOT alter or change valve assembly. - Risk of Injury To reduce the risk of oil and grease to follow these safety instructions..., Injury or Damage from service and condemned by trained and authorized personnel. Contact your dealer/Home Care equipment provider for service. • Dents or ...
Owners Manual
Page 17
...to the HomeFill Compressor user manual for visible damage. If below 1500 psig (10342 kPa), then refill the cylinder assembly using the HomeFill system. ... use , inspect product for operating instructions to connect and disconnect the cylinder to fill. - Prior to refill the cylinder assembly on the Invacare HomeFill...Invacare dealer or qualified technician for any reason, any damage is recommended to each use near any type of flame or flammable/explosive substances, vapors or atmosphere. 1112648-H-00 Usage 1. If for service or replacement. 4 Usage 4.1 Before Use WARNING! DO NOT use...
...to the HomeFill Compressor user manual for visible damage. If below 1500 psig (10342 kPa), then refill the cylinder assembly using the HomeFill system. ... use , inspect product for operating instructions to connect and disconnect the cylinder to fill. - Prior to refill the cylinder assembly on the Invacare HomeFill...Invacare dealer or qualified technician for any reason, any damage is recommended to each use near any type of flame or flammable/explosive substances, vapors or atmosphere. 1112648-H-00 Usage 1. If for service or replacement. 4 Usage 4.1 Before Use WARNING! DO NOT use...
Owners Manual
Page 18
... manufacturer. 8. To avoid injury to patient, ALWAYS confirm prescribed setting before administering to setting "Off". 18 1112648-H-00 Breathe through the nose and feel a flow of oxygen as instructed by your nose as you inhale. 11. Invacare® ML4, ML6 and M9 Cylinder with a maximum length ... instructions. Position the cylinder assembly A so the outlet connection D is pointing away from the user and any person, or flammable material when adjusting the flow dial E. 4. The cannula MUST be routed and secured properly. - To avoid injury or death: - When finished using ...
... manufacturer. 8. To avoid injury to patient, ALWAYS confirm prescribed setting before administering to setting "Off". 18 1112648-H-00 Breathe through the nose and feel a flow of oxygen as instructed by your nose as you inhale. 11. Invacare® ML4, ML6 and M9 Cylinder with a maximum length ... instructions. Position the cylinder assembly A so the outlet connection D is pointing away from the user and any person, or flammable material when adjusting the flow dial E. 4. The cannula MUST be routed and secured properly. - To avoid injury or death: - When finished using ...
Owners Manual
Page 20
...Ensure the cover is pointing away from the user and any other person(s). 2. DO NOT ...setting "5". 4. Follow the cleaning/disinfectant product manufacturer's instructions for use , handling, storage and disposal of Damage - Follow this procedure. Clean exterior surfaces of the product. Position the cylinder assembly A so the outlet connection D is installed...instructions for use , handling, storage and disposal of the cylinder with a lint free cloth dampened with Conserver CAUTION! If the conserver is dry before performing this procedure to setting "Off". 3. Invacare...
...Ensure the cover is pointing away from the user and any other person(s). 2. DO NOT ...setting "5". 4. Follow the cleaning/disinfectant product manufacturer's instructions for use , handling, storage and disposal of Damage - Follow this procedure. Clean exterior surfaces of the product. Position the cylinder assembly A so the outlet connection D is installed...instructions for use , handling, storage and disposal of the cylinder with a lint free cloth dampened with Conserver CAUTION! If the conserver is dry before performing this procedure to setting "Off". 3. Invacare...
Owners Manual
Page 21
... free cloth. Invacare products are specifically designed and manufactured for use with Invacare accessories. Accessories designed by other sources of contamination. 5.7 Optional Accessories WARNING! To obtain Invacare accessories, contact Invacare by Invacare and are available for use in conjunction with Invacare products. - The following procedures: - DO NOT wipe the fillport or outlet connection. 4. Before transferring the product to setting "Off...
... free cloth. Invacare products are specifically designed and manufactured for use with Invacare accessories. Accessories designed by other sources of contamination. 5.7 Optional Accessories WARNING! To obtain Invacare accessories, contact Invacare by Invacare and are available for use in conjunction with Invacare products. - The following procedures: - DO NOT wipe the fillport or outlet connection. 4. Before transferring the product to setting "Off...
Owners Manual
Page 22
Invacare® ML4, ML6 and M9 Cylinder with Conserver 6 After Use 6.1 Disposal This product has been supplied from an environmentally aware manufacturer that complies with the product for specific warranty information. Follow local governing ordinances and recycling plans regarding disposal ...of in places (landfills) that are not appropriate according to legislation. This product may contain substances that could be handled in the normal waste stream. • Any accessories not part...
Invacare® ML4, ML6 and M9 Cylinder with Conserver 6 After Use 6.1 Disposal This product has been supplied from an environmentally aware manufacturer that complies with the product for specific warranty information. Follow local governing ordinances and recycling plans regarding disposal ...of in places (landfills) that are not appropriate according to legislation. This product may contain substances that could be handled in the normal waste stream. • Any accessories not part...
Owners Manual
Page 23
... points to see if the cannula is between settings. 4. Check to a setting. 4. Possible Cause 1. Cylinder empty. 3. Kinked or blocked cannula. Solution 1. Fill or replace cylinder. 3. Flow dial is kinked. 1112648-H-00 23 7 Troubleshooting Troubleshooting 7.1 Troubleshooting If the Regulating Cylinder fails to prescribed setting. 2. Adjust flow dial to function, consult the Troubleshooting guide. Problem No flow. Check position of cannula in...
... points to see if the cannula is between settings. 4. Check to a setting. 4. Possible Cause 1. Cylinder empty. 3. Kinked or blocked cannula. Solution 1. Fill or replace cylinder. 3. Flow dial is kinked. 1112648-H-00 23 7 Troubleshooting Troubleshooting 7.1 Troubleshooting If the Regulating Cylinder fails to prescribed setting. 2. Adjust flow dial to function, consult the Troubleshooting guide. Problem No flow. Check position of cannula in...
Owners Manual
Page 25
... to 140°F (-40°C to 60°C) 95% RH Assembly Dimensions Conserver S/N Range: Model: HF2PCL4 Weight: N/A Overall Length: N/A Diameter: N/A 1-69,999 75,000+ HF2PCL6 4.3 lb (1.95 kg) 11.1 in (28 cm) 4.4 in (11.1 cm) HF2PC9 5.3 lb (2.4 kg) 14.2 in (36 cm) 4.4 in (11.1 cm) HF2PCL4 3.6 lbs... (1.63 kg) 8.9 in (22.2 cm) 4.4 in (11.1 cm) HF2PCL6 4.1 lb (1.86 kg) 10.9 in (27.7 cm) 4.4 in (11.1 cm) HF2PC9 5.1 lb (2.31 kg) 14 in (35.5 cm) 4.4 in (11.1 cm) Cylinders Cylinder Size: Volume (H2O): Oxygen Capacity: ML4 0.67 L (0.024 ft 3) 113 L (4 ft 3) ...
... to 140°F (-40°C to 60°C) 95% RH Assembly Dimensions Conserver S/N Range: Model: HF2PCL4 Weight: N/A Overall Length: N/A Diameter: N/A 1-69,999 75,000+ HF2PCL6 4.3 lb (1.95 kg) 11.1 in (28 cm) 4.4 in (11.1 cm) HF2PC9 5.3 lb (2.4 kg) 14.2 in (36 cm) 4.4 in (11.1 cm) HF2PCL4 3.6 lbs... (1.63 kg) 8.9 in (22.2 cm) 4.4 in (11.1 cm) HF2PCL6 4.1 lb (1.86 kg) 10.9 in (27.7 cm) 4.4 in (11.1 cm) HF2PC9 5.1 lb (2.31 kg) 14 in (35.5 cm) 4.4 in (11.1 cm) Cylinders Cylinder Size: Volume (H2O): Oxygen Capacity: ML4 0.67 L (0.024 ft 3) 113 L (4 ft 3) ...
Owners Manual
Page 26
Invacare® ML4, ML6 and M9 Cylinder with Conserver HF2PCL6 HF2PC9 *With Concentrator flow rates to patient: Concentrator Model Platinum 5 and Perfecto2 Platinum 10 1 hr 25 min* 2 hr 20 min* Concentrator Flow Rate Up to 3 L/min Up to 5 ... on environmental conditions. 1 hr 50 min* 2 hr 45 min* Accessories Model Single Lumen Cannula Carry Bag HF2PCL4 HF2PCL6 HF2PC9 4' (1.21 m) recommended. 7' (2.1 m) tubing length maximum HF2PCL4 Carry Bag HF2PCL6 Carry Bag HF2PC9 Carry Bag Regulatory Compliance CGA G-4.1 (Cleaning Equipment For Oxygen Service) DOT & TC (3AL & 3ALM) 26 1112648-H-00
Invacare® ML4, ML6 and M9 Cylinder with Conserver HF2PCL6 HF2PC9 *With Concentrator flow rates to patient: Concentrator Model Platinum 5 and Perfecto2 Platinum 10 1 hr 25 min* 2 hr 20 min* Concentrator Flow Rate Up to 3 L/min Up to 5 ... on environmental conditions. 1 hr 50 min* 2 hr 45 min* Accessories Model Single Lumen Cannula Carry Bag HF2PCL4 HF2PCL6 HF2PC9 4' (1.21 m) recommended. 7' (2.1 m) tubing length maximum HF2PCL4 Carry Bag HF2PCL6 Carry Bag HF2PC9 Carry Bag Regulatory Compliance CGA G-4.1 (Cleaning Equipment For Oxygen Service) DOT & TC (3AL & 3ALM) 26 1112648-H-00