Owners Manual
Page 1
User: Before using this product, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare.com Operator's Manual Platinum™ Series XL, 5, 10 Oxygen Concentrators Standard, with SensO2®and HomeFill® Invacare 5 with SensO2 Model IRC5LXO2 Invacare 10 with SensO2 Model IRC10LXO2 Invacare 5 Model IRC5LX Invacare 10 Model IRC10LX Dealer: This manual MUST be given to the end user.
User: Before using this product, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.invacare.com Operator's Manual Platinum™ Series XL, 5, 10 Oxygen Concentrators Standard, with SensO2®and HomeFill® Invacare 5 with SensO2 Model IRC5LXO2 Invacare 10 with SensO2 Model IRC10LXO2 Invacare 5 Model IRC5LX Invacare 10 Model IRC10LX Dealer: This manual MUST be given to the end user.
Owners Manual
Page 8
...the safe operation and use any lubricants unless recommended by Invacare. DO NOT use of this product. SECTION 1-GENERAL GUIDELINES SECTION 1-GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the Platinum concentrator these instructions MUST be followed. ƽ WARNING This ...Refer servicing to Persons Ꮨ DANGER Risk of friction. DO NOT come in another room at least 7 ft from the oxygen concentrator, tubing and connections, and all other liquid. DO NOT place or store product where it is required by any spark near medical...
...the safe operation and use any lubricants unless recommended by Invacare. DO NOT use of this product. SECTION 1-GENERAL GUIDELINES SECTION 1-GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the Platinum concentrator these instructions MUST be followed. ƽ WARNING This ...Refer servicing to Persons Ꮨ DANGER Risk of friction. DO NOT come in another room at least 7 ft from the oxygen concentrator, tubing and connections, and all other liquid. DO NOT place or store product where it is required by any spark near medical...
Owners Manual
Page 9
... IEC/EN 60601‐1‐2. Radio Frequency Interference This equipment has been tested and found to comply with other oxygen concentrators or oxygen therapy devices. Close supervision is necessary when this product is Off when not in parallel or series with EMC ... Keep the oxygen tubing, cord, and unit out from under such items as blankets, bed coverings, chair cushions, clothing and away from heated or hot surfaces, including space heaters, stoves and similar electrical appliances. These limits are no user serviceable parts. Invacare recommends that each concentrator be left...
... IEC/EN 60601‐1‐2. Radio Frequency Interference This equipment has been tested and found to comply with other oxygen concentrators or oxygen therapy devices. Close supervision is necessary when this product is Off when not in parallel or series with EMC ... Keep the oxygen tubing, cord, and unit out from under such items as blankets, bed coverings, chair cushions, clothing and away from heated or hot surfaces, including space heaters, stoves and similar electrical appliances. These limits are no user serviceable parts. Invacare recommends that each concentrator be left...
Owners Manual
Page 11
... - The outlet fitting does not affect concentrator performance. For more information about the HomeFill, contact your Invacare dealer. IOH200 Power Cord Outlet Fitting NOTE: This outlet fitting is to the HomeFill ownerʹs manual, part number 1100873, for filling oxygen cylinders with the HomeFill home oxygen compressor. When not in use, the plug provided...
... - The outlet fitting does not affect concentrator performance. For more information about the HomeFill, contact your Invacare dealer. IOH200 Power Cord Outlet Fitting NOTE: This outlet fitting is to the HomeFill ownerʹs manual, part number 1100873, for filling oxygen cylinders with the HomeFill home oxygen compressor. When not in use, the plug provided...
Owners Manual
Page 12
... dry area. 2. Inspection Inspect/examine exterior of repackaged concentrator. Remove all components. DO NOT place objects on top of the oxygen concentrator for nicks, dents, scratches or other damages. Unpacking NOTE...oxygen concentrator in the upright position to FIGURE 3.1. 1. SECTION 3-HANDLING SECTION 3-HANDLING The concentrator should ALWAYS be reshipped by common carrier, additional cartons are available from the carton. Storage 1. If the concentrator is evident, notify the carrier, or your local dealer. 2. Carefully remove all the components from Invacare...
... dry area. 2. Inspection Inspect/examine exterior of repackaged concentrator. Remove all components. DO NOT place objects on top of the oxygen concentrator for nicks, dents, scratches or other damages. Unpacking NOTE...oxygen concentrator in the upright position to FIGURE 3.1. 1. SECTION 3-HANDLING SECTION 3-HANDLING The concentrator should ALWAYS be reshipped by common carrier, additional cartons are available from the carton. Storage 1. If the concentrator is evident, notify the carrier, or your local dealer. 2. Carefully remove all the components from Invacare...
Owners Manual
Page 13
... XL: 5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa) Platinum 10: 9 psi ± 0.5 psi (62.0 kPa ± 3.45 kPa) *NOTE: Stated concentration levels achieved after initial warm‐up period (approximately 30 minutes). From 4,000 ft (1230 meters) to 13,129 ft (4000 meters) below 90% efficiency... XL: 50 dB Average Platinum 10: 58 dB Average Altitude: Platinum 5 and XL - Up to 13,129 ft (4000 meters) below 90% efficiency *Oxygen Output Concentration Levels: All 5LXO2/5LX models 87% to 95.6% for flows ranging 0.5 to 5 L/min and all 10LXO2/10LX models 87% to 95.6% for flows ranging...
... XL: 5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa) Platinum 10: 9 psi ± 0.5 psi (62.0 kPa ± 3.45 kPa) *NOTE: Stated concentration levels achieved after initial warm‐up period (approximately 30 minutes). From 4,000 ft (1230 meters) to 13,129 ft (4000 meters) below 90% efficiency... XL: 50 dB Average Platinum 10: 58 dB Average Altitude: Platinum 5 and XL - Up to 13,129 ft (4000 meters) below 90% efficiency *Oxygen Output Concentration Levels: All 5LXO2/5LX models 87% to 95.6% for flows ranging 0.5 to 5 L/min and all 10LXO2/10LX models 87% to 95.6% for flows ranging...
Owners Manual
Page 16
...NOT place in a well ventilated area to you use your concentrator and will serve as a reference as a bed or couch, where the air opening may select a room where using your oxygen concentrator. The air intake of oxygen delivery systems. Your provider will tell you how to other types... of the unit should be contacted with any questions or problems regarding your oxygen concentrator would be blocked. This owner's manual will ...
...NOT place in a well ventilated area to you use your concentrator and will serve as a reference as a bed or couch, where the air opening may select a room where using your oxygen concentrator. The air intake of oxygen delivery systems. Your provider will tell you how to other types... of the unit should be contacted with any questions or problems regarding your oxygen concentrator would be blocked. This owner's manual will ...
Owners Manual
Page 19
... the unit is flowing through the cannula. Attach the cannula/patient supply tubing to FIGURE 5.5 on the oxygen concentrator (FIGURE 5.4). 8. Oxygen Tubing Oxygen Outlet Connector Humidifier Outlet FIGURE 5.4 Attaching the Oxygen Tubing Power Switch NOTE: For this procedure, refer to the humidifier bottle outlet (FIGURE 5.4). 9. SECTION .../Off (I/O) Power Switch FIGURE 5.5 Power Switch Part No 1118353 19 Platinum™ Series Press power switch to oxygen outlet connector on page 19. 1. Attach oxygen tubing from the humidifier bottle to on for one second, indicating that...
... the unit is flowing through the cannula. Attach the cannula/patient supply tubing to FIGURE 5.5 on the oxygen concentrator (FIGURE 5.4). 8. Oxygen Tubing Oxygen Outlet Connector Humidifier Outlet FIGURE 5.4 Attaching the Oxygen Tubing Power Switch NOTE: For this procedure, refer to the humidifier bottle outlet (FIGURE 5.4). 9. SECTION .../Off (I/O) Power Switch FIGURE 5.5 Power Switch Part No 1118353 19 Platinum™ Series Press power switch to oxygen outlet connector on page 19. 1. Attach oxygen tubing from the humidifier bottle to on for one second, indicating that...
Owners Manual
Page 21
... For this procedure, refer to FIGURE 5.7. Have a back-up supply of the Concentrator NOTE: Concentrator may be used during the initial start warm‐up supply of oxygen generated by the oxygen concentrator. Switch to a back-up time (approximately 30 min.) while waiting for the ...Indicators - YELLOW Flashing Sensor Failure Call a qualified technician. Part No 1118353 21 Platinum™ Series OXYGEN PURITY Auto Shut - Units with SensO2 Initial Startup of oxygen ready. YELLOW Solid B. Call a qualified technician. If purity falls below factory preset standards, indicator lights...
... For this procedure, refer to FIGURE 5.7. Have a back-up supply of the Concentrator NOTE: Concentrator may be used during the initial start warm‐up supply of oxygen generated by the oxygen concentrator. Switch to a back-up time (approximately 30 min.) while waiting for the ...Indicators - YELLOW Flashing Sensor Failure Call a qualified technician. Part No 1118353 21 Platinum™ Series OXYGEN PURITY Auto Shut - Units with SensO2 Initial Startup of oxygen ready. YELLOW Solid B. Call a qualified technician. If purity falls below factory preset standards, indicator lights...
Owners Manual
Page 22
... technician. O2 purity will build to a back-up time (approximately 30 min). Switch to a maximum during the initial start warm‐up supply of the Concentrator NOTE: Concentrator may be used during this procedure, refer to FIGURE 5.8. LABEL SYMBOL STATUS INDICATOR LIGHTS (LED) I /O GREEN FIGURE 5.8 Front Panel Indicators - Units without SensO2 NOTE...
... technician. O2 purity will build to a back-up time (approximately 30 min). Switch to a maximum during the initial start warm‐up supply of the Concentrator NOTE: Concentrator may be used during this procedure, refer to FIGURE 5.8. LABEL SYMBOL STATUS INDICATOR LIGHTS (LED) I /O GREEN FIGURE 5.8 Front Panel Indicators - Units without SensO2 NOTE...
Owners Manual
Page 24
...rinse thoroughly. 3. Cleaning the Humidifier NOTE: To clean the oxygen humidifier, follow these STEPS: 1. To avoid electrical shock, DO NOT remove cabinet. Clean the cabinet filters with a solution of concentrator) FIGURE 6.1 Cleaning the Cabinet Filter Cleaning the Cabinet ƽ...; WARNING UNPLUG the concentrator when cleaning. SECTION 6-MAINTENANCE 2. If none are provided, follow the instructions provided by...
...rinse thoroughly. 3. Cleaning the Humidifier NOTE: To clean the oxygen humidifier, follow these STEPS: 1. To avoid electrical shock, DO NOT remove cabinet. Clean the cabinet filters with a solution of concentrator) FIGURE 6.1 Cleaning the Cabinet Filter Cleaning the Cabinet ƽ...; WARNING UNPLUG the concentrator when cleaning. SECTION 6-MAINTENANCE 2. If none are provided, follow the instructions provided by...
Owners Manual
Page 25
...L/min. SECTION 6-MAINTENANCE SERIAL NO. Flow Rate EVERY 4,380 HOURS FOR IRC5LX (EVERY 2,160 HOURS FOR IRC10LX) Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace...7 days per week, for 3 years. I RC Part No 1118353 FIGURE 6.2 Preventive Maintenance Record 25 PREVENTIVE MAINTENANCE RECORD INVACARE PLATINUM OXYGEN CONCENTRATOR ON EACH INSPECTION Record Date of Service Manual. Platinum™ Series MODEL NO.
...L/min. SECTION 6-MAINTENANCE SERIAL NO. Flow Rate EVERY 4,380 HOURS FOR IRC5LX (EVERY 2,160 HOURS FOR IRC10LX) Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace...7 days per week, for 3 years. I RC Part No 1118353 FIGURE 6.2 Preventive Maintenance Record 25 PREVENTIVE MAINTENANCE RECORD INVACARE PLATINUM OXYGEN CONCENTRATOR ON EACH INSPECTION Record Date of Service Manual. Platinum™ Series MODEL NO.
Owners Manual
Page 26
...) for kinks or 1. Inspect house circuit break- Beep...Beep Low Flow Alarm: 1. Concentrator ers or fuses. outlet. Concentrator NOT operating, Power Switch ON. 2. cords. Inspect for Platinum 10. Platinum™ Series 26 Part No 1118353 Call Service Provider. Move oxygen concentrator at 1b. Power cord NOT plugged in. 1. No power at outlet. 2. blocked air...
...) for kinks or 1. Inspect house circuit break- Beep...Beep Low Flow Alarm: 1. Concentrator ers or fuses. outlet. Concentrator NOT operating, Power Switch ON. 2. cords. Inspect for Platinum 10. Platinum™ Series 26 Part No 1118353 Call Service Provider. Move oxygen concentrator at 1b. Power cord NOT plugged in. 1. No power at outlet. 2. blocked air...
Owners Manual
Page 27
...cabinet blocked air intake. GREEN LIGHT with YELLOW Light Flashing. 1. Correct, clean or replace Alarm: item. Check flowmeter is set too low: 3. Move concentrator at outlet. 5. DO NOT use extension cords. Internal repairs required. 1. Part No 1118353 27 Platinum™ Series SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE ...1. filters. 4b. Internal repairs required. 6. Clean or Replace filters. then turn CONTINUOUS power OFF for 1.0 L/min. Call Service Provider. Low oxygen purity.* 1. ing, cannula or humidifier.* ages. Kinked or blocked tub- 2.
...cabinet blocked air intake. GREEN LIGHT with YELLOW Light Flashing. 1. Correct, clean or replace Alarm: item. Check flowmeter is set too low: 3. Move concentrator at outlet. 5. DO NOT use extension cords. Internal repairs required. 1. Part No 1118353 27 Platinum™ Series SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE ...1. filters. 4b. Internal repairs required. 6. Clean or Replace filters. then turn CONTINUOUS power OFF for 1.0 L/min. Call Service Provider. Low oxygen purity.* 1. ing, cannula or humidifier.* ages. Kinked or blocked tub- 2.
Owners Manual
Page 30
...invoice required for coverage under this warranty shall be limited to the original purchaser who purchases this warranty. Invacare Corporation warrants its Platinum XL Oxygen Concentrators when purchased new and unused to the valve body. This will end upon any such subsequent sale or ...all warranty applicable to be repaired or replaced, at Invacare Corporation's option. This warranty gives you specific legal rights and you may vary from Invacare Corporation or a dealer. Invacare Corporation warrants its Platinum 10 Oxygen Concentrators when purchased new and unused to state. This ...
...invoice required for coverage under this warranty shall be limited to the original purchaser who purchases this warranty. Invacare Corporation warrants its Platinum XL Oxygen Concentrators when purchased new and unused to the valve body. This will end upon any such subsequent sale or ...all warranty applicable to be repaired or replaced, at Invacare Corporation's option. This warranty gives you specific legal rights and you may vary from Invacare Corporation or a dealer. Invacare Corporation warrants its Platinum 10 Oxygen Concentrators when purchased new and unused to state. This ...