CS616 and CS625 Water Content Reflectometers
Page 1
INSTRUCTION MANUAL CS616 and CS625 Water Content Reflectometers Revision: 2/14 Copyright © 2002-2014 Campbell Scientific, Inc.
INSTRUCTION MANUAL CS616 and CS625 Water Content Reflectometers Revision: 2/14 Copyright © 2002-2014 Campbell Scientific, Inc.
CS616 and CS625 Water Content Reflectometers
Page 3
...apply to products manufactured by the original manufacturer. The customer shall assume all other locations, Campbell will return such products by Campbell Scientific, Inc. ("Campbell") to Campbell. Campbell will return such products best way CIP (Port of removing, reinstalling, and shipping defective ... materials and workmanship under this warranty is in the corresponding Campbell pricelist or product manual. Campbell is part of America. Products not manufactured, but that are re-sold by Campbell, are warranted by surface carrier prepaid within the continental United...
...apply to products manufactured by the original manufacturer. The customer shall assume all other locations, Campbell will return such products by Campbell Scientific, Inc. ("Campbell") to Campbell. Campbell will return such products best way CIP (Port of removing, reinstalling, and shipping defective ... materials and workmanship under this warranty is in the corresponding Campbell pricelist or product manual. Campbell is part of America. Products not manufactured, but that are re-sold by Campbell, are warranted by surface carrier prepaid within the continental United...
CS616 and CS625 Water Content Reflectometers
Page 14
...'Declare Public and Dim Variables Public batt_volt Public Panel_temp Public Period (8) Public VWC (8) Public Flag (1) Dim I 'Declare Constants 'CS616 Default Calibration Constants const a0= -0.0663 const a1= -0.0063 const a2= 0.0007 'Flag logic constants const high = true const ... manually set flag 1 high/low) If IfTime (0,30,min) Then flag (1) = high If Flag (1) = high Then 'measure 8ea CS616 probes on CR1000 CS616 (Period(1),4,9,7,4,1.0,0) 'measure 4ea CS616 probes, enable w/ C7 CS616 (Period(5),4,13,8,4,1.0,0) 'measure 4ea CS616 probes, enable w/ C8 ' For I=1 to 8 'convert CS616 ...
...'Declare Public and Dim Variables Public batt_volt Public Panel_temp Public Period (8) Public VWC (8) Public Flag (1) Dim I 'Declare Constants 'CS616 Default Calibration Constants const a0= -0.0663 const a1= -0.0063 const a2= 0.0007 'Flag logic constants const high = true const ... manually set flag 1 high/low) If IfTime (0,30,min) Then flag (1) = high If Flag (1) = high Then 'measure 8ea CS616 probes on CR1000 CS616 (Period(1),4,9,7,4,1.0,0) 'measure 4ea CS616 probes, enable w/ C7 CS616 (Period(5),4,13,8,4,1.0,0) 'measure 4ea CS616 probes, enable w/ C8 ' For I=1 to 8 'convert CS616 ...
CS616 and CS625 Water Content Reflectometers
Page 16
... this channel number. SEChan: The SEChan argument is used to measure the period (in microseconds) or the frequency (in kHz) of the measurement. CS616 and CS625 Water Content Reflectometers 'Main Program BeginProg Scan (5,Sec,0,0) 'scan instructions every 5 sec Battery (Batt_volt) PanelTemp (Panel_temp,250) ' 'Set... the single-ended channel on which to make the measurement. This instruction can manually set flag 1 high/low) If IfTime (0,30,min) Then flag (1) = high If Flag (1) = high Then 'measure 48ea CS616 probes on AM16/32 in which to store the results of a signal on...
... this channel number. SEChan: The SEChan argument is used to measure the period (in microseconds) or the frequency (in kHz) of the measurement. CS616 and CS625 Water Content Reflectometers 'Main Program BeginProg Scan (5,Sec,0,0) 'scan instructions every 5 sec Battery (Batt_volt) PanelTemp (Panel_temp,250) ' 'Set... the single-ended channel on which to make the measurement. This instruction can manually set flag 1 high/low) If IfTime (0,30,min) Then flag (1) = high If Flag (1) = high Then 'measure 48ea CS616 probes on AM16/32 in which to store the results of a signal on...
CS616 and CS625 Water Content Reflectometers
Page 27
Can execute measurement sequence manually by setting Flag 1 high. 21 The AM16/32-series multiplexer slide switch on the top panel is set to datalogger final 13: End (P95) *Table 2 Program 02: 0.0000 Execution Interval (seconds) *Table 3 Subroutines End Program -Input Locations- 1 TSoil 1 1 1 2 Tref 1 1 1 3 CS616 1 2 1 4 TempCS616 1 2 2 5 TFactor 1 1 1 6 NewCS616 1 1 1 7 WaterCont 1 1 1 6.4.5.5 CS616 Sample Edlog Program 5 Hourly...
Can execute measurement sequence manually by setting Flag 1 high. 21 The AM16/32-series multiplexer slide switch on the top panel is set to datalogger final 13: End (P95) *Table 2 Program 02: 0.0000 Execution Interval (seconds) *Table 3 Subroutines End Program -Input Locations- 1 TSoil 1 1 1 2 Tref 1 1 1 3 CS616 1 2 1 4 TempCS616 1 2 2 5 TFactor 1 1 1 6 NewCS616 1 1 1 7 WaterCont 1 1 1 6.4.5.5 CS616 Sample Edlog Program 5 Hourly...